Operation
Configure a 4-20 mA input module
An analog module must be installed in the controller.
1.
Determine what output the the connected device is using (0-20 mA or
4-20 mA). This information will be used to set the scale.
2.
Determine what the 20 mA value is equal to (e.g, 100 psi).
3.
Determine what the low end (0 or 4 mA) value is equal to (e.g., 10 psi).
This information will be used to set the display range.
4.
From the Settings Menu, select Sensor Setup.
5.
Select Configure.
6.
Update the options.
a.
Highlight an option and push
ENTER
.
b.
Make a selection or update the entries.
c.
Push
ENTER
to save the changes.
Option
Description
Edit name
Edits the module name
Edit units
Edits the measurement units
Edit parameter
Edits the parameter name
Display range
For the 0-20 mA scale:
• Set the 20 mA value
• Set the 0 mA value
For the 4-20 mA scale:
• Set the 20 mA value
• Set the 4 mA value
Sets the values used for the selected
scale (0-20 mA or 4-20 mA)
Option
Description
Signal average
Sets how often signals are averaged.
Higher values produce a smoother
signal but increase the time it takes
for a signal to respond to a change in
the process value.
Set resolution—
X.XXX, XX.XX, XXX.X, XXXX
Sets the number of decimal places
used in the display.
Select scale—
4-20 mA or 0-20 mA
Sets scale used for the 4-20 mA input
Data log interval—5 sec, 30 sec, 1
min, 2 min, 5 min, 10 min, 15 min, 30
min, 60 min
Sets how often data is logged to the
internal controller memory.
Reset defaults—Push
ENTER
to
reset configuration settings or push
the
BACK
key to cancel.
Resets configuration settings to the
default values.
Calibration
Notes
• Output Mode
• Set the output mode to Active to use the current measured value in
a calibration.
• Set the output mode to Hold to use the last measured value in a
calibration.
• Set the output mode to Transfer to transfer a specified value to the
4-20 mA input for use in a calibration.
• Cal Options
• Select Current to perform a calibration using current values (mA).
• Select Input to perform a calibration using measurement values (such
as psi).
English
7
Содержание sc200 4-20
Страница 2: ...English 3 Fran ais 10 Espa ol 18 Portugu s 25 32 39 46 53 2...
Страница 5: ...1 2 3 4 English 5...
Страница 6: ...5 6 7 8 6 English...
Страница 12: ...1 2 3 4 12 Fran ais...
Страница 13: ...5 6 7 8 Fran ais 13...
Страница 20: ...1 2 3 4 20 Espa ol...
Страница 21: ...5 6 7 8 Espa ol 21...
Страница 27: ...1 2 3 4 Portugu s 27...
Страница 28: ...5 6 7 8 28 Portugu s...
Страница 32: ...2005 8 12 2002 98 EC 4 20 mA 4 20 mA 0 20 mA 4 20 mA Modbus Modbus www hach com www hach lange com 32...
Страница 33: ...4 20 mA 33...
Страница 34: ...1 2 3 4 34...
Страница 35: ...5 6 7 8 35...
Страница 39: ...Web 2005 8 12 EU 2002 98 EC 4 20mA 4 20mA 0 20 mA 4 20 mA 1 Modbus Modbus www hach com www hach lange com 39...
Страница 40: ...4 20mA 40...
Страница 41: ...1 2 3 4 41...
Страница 42: ...5 6 7 8 42...
Страница 46: ...2005 8 12 EU 2002 98 EC 4 20 mA 4 20 mA 0 20 mA 4 20 mA Modbus Modbus www hach com www hach lange com 46...
Страница 47: ...4 20 mA 47...
Страница 48: ...1 2 3 4 48...
Страница 49: ...5 6 7 8 49...
Страница 54: ...4 20 mA 54...
Страница 55: ...1 2 3 4 55...
Страница 56: ...5 6 7 8 56...
Страница 60: ...www hach com Hach Company 2010 All rights reserved Printed in the U S A...