5.
Select an output mode option and push
enter
.
Option
Description
ACTIVE
Output level continues to represent operating conditions.
HOLD
Output level is held static.
TRANSFER
Output level moves to a pre-configured value.
6.
Select the sensors that will have their outputs (analog and relay) affected when one of the
discrete inputs becomes active. Push
enter
.
7.
Use the arrows to select the ON DELAY value (the duration time delay between the discrete input
activation and the configured response of the controller). Push
enter
.
8.
Use the arrows to select the OFF DELAY value (the duration time delay between the discrete
input de-activation and the configured response of the controller). Push
enter
.
9.
Repeat steps 4–8 for each desired channel.
10.
If a discrete input needs to be changed after initially set up:
a.
Repeat steps 1–2 and an INPUT SETTINGS menu appears with the following options:
• CONTROL LOGIC
• SET WARNING
• OUTPUT MODE
• ON DELAY
• OFF DELAY
b.
Select the desired option and push
enter
.
c.
Make the desired changes and push
enter
to save the changes and return to the INPUT
SETTINGS menu.
Set up a calculation
Two sensors must be installed in the controller.
1.
Push the
menu
key and select Polymetron 9500 SETUP>CALCULATION.
2.
Select a menu option and choose from the displayed list or update the entry.
Option
Description
SET VARIABLE X
Selects the sensor that corresponds to the variable set as "X".
SET PARAMETER X
Selects the measurement type that corresponds to the variable set as "X".
SET VARIABLE Y
Selects the sensor that corresponds to the variable set as "Y".
SET PARAMETER Y
Selects the measurement type that corresponds to the variable set as "Y".
SET FORMULA
NONE, X-Y, X/Y, [X/Y]%, [X+Y]/2, X*Y, [X-Y]%/X.
DISPLAY FORMAT
Sets the number of decimal places displayed in a calculation result.
SET UNITS
Sets the units of measurement.
SET PARAMETER
Assigns the name of the measurement.
3.
Push
enter
to save the selection or setting and return to the CALCULATION menu.