Advanced operation
Security setup
Enable or disable the pass code
By default the pass code option is disabled and all configuration settings and calibrations can be
changed. When the pass code function is enabled, access to a number of menus and menu options
will require a pass code.
1.
Push the
menu
key and select Polymetron 9500 SETUP>SECURITY SETUP>SET PASS CODE.
2.
Select DISABLED or ENABLED and push
enter
.
3.
Push the
home
key to enable the new security setting.
Edit the pass code
The pass code is factory set to sc200. The EDIT PASS CODE menu option appears in the
SECURITY SETUP menu only after the pass code feature is enabled and a valid pass code has
been entered.
A pass code consists of up to six upper or lower-case alpha, numeric and special characters. Pass
codes are case-sensitive.
To edit the pass code:
1.
Make sure the pass code is enabled. Refer to
Enable or disable the pass code
on page 24 for
information on how to enable the pass code.
2.
Push the
menu
key and select Polymetron 9500 SETUP>SECURITY SETUP.
3.
Use the arrow keys to enter the current valid pass code and push
enter
.
4.
Select EDIT PASS CODE and push
enter
.
5.
Use the arrow keys to edit the pass code and push
enter
.
6.
Push the
home
key to save the new pass code.
Configure a 4-20 mA input module
An analog module must be installed in the controller.
1.
Determine what output the connected device is using (0-20 mA or 4-20 mA). This information will
be used to set the scale.
2.
Determine what the 20 mA value is equal to (e.g. 100 psi).
3.
Determine what the low end (0 or 4 mA) value is equal to (e.g. 10 psi). This information will be
used to set the display range.
4.
Push the
menu
key and select SENSOR SETUP>CONFIGURE.
5.
Update the options.
a.
Highlight an option and push
enter
.
b.
Make a selection or update the entries.
c.
Push
enter
to save the changes.
Option
Description
EDIT NAME
Edits the module name
EDIT UNITS
Edits the measurement units
EDIT PARAMETER
Edits the parameter name
24
English
Содержание Polymetron 9500
Страница 1: ...DOC023 52 93058 Polymetron 9500 Controller 05 2015 Edition 6 User Manual...
Страница 2: ......
Страница 10: ...Controller mounting Figure 3 Surface mounting dimensions 8 English...
Страница 12: ...Figure 5 Pipe mounting vertical pipe 10 English...
Страница 49: ......