3.
Press
Select
to move from one field to another to keep the value displayed, or press
Enter
on
completion of a field to move to the next.
4.
Press the
Esc
function key on completion.
Priming tubes
1.
After refilling the bottles (reference electrolyte, calibration solution and reactivation solution)
and/or after exchanging the empty bottle of conditioning reagent with a new one, select the
PRIMING TUBES
option. This will set off the process to purge and refill the tubes.
2.
Press the
Esc
function key at any time to abort the process and return to the menu.
Bottles full
1.
If the bottles of reagent are full, select
Yes
and the default values for the volumes will be set.
These values can be viewed and modified by selecting the
ADJUST BOTTLE VOLS.
option in
the menu. If you select
No
to this option, you will manually have to enter the correct values using
the
ADJUST BOTTLE VOLS.
option.
Test accessories
This option allows you to check that a number of accessories, as well as the alarm relays and logical
inputs are functioning correctly.
Sensor reactivation
Usually, the electrode is reactivated automatically based on the time delay set in
on page 16. This option allows for a manual reactivation.
Extended stop
If the instrument is not to be used for an extended period of time, select this option to shut the
analyzer down in a controlled manner.
1.
Place all the tubes in demineralized water for cleaning.
2.
Press
OK
to continue.
3.
A progress bar will be displayed whilst the shut down is being performed.
4.
On completion, you will be informed to remove and store the electrodes.
5.
Press
ESC
to exit.
Startup
This process guides you through various steps required to set the instrument up for initial
measurements, or to restart after a long period of inactivity. This includes regulating the sample flow
and purging the pumps of reagent.
1.
Regulate the flow of the sample to ensure a slight overflow into the overflow vessel.
2.
Select
OK
to continue. You will then be required to regulate the flow on the next configured
channel.
3.
On completion, an
Action completed
message will be displayed. Select
Esc
to exit.