![Hach POLYMETRON 9135 Скачать руководство пользователя страница 45](http://html.mh-extra.com/html/hach/polymetron-9135/polymetron-9135_operators-manual_577103045.webp)
Transmitter 9135 – pH/redox measurement
Page 41
TEMP. COMP.
This menu allows to set up the temperature
compensation.
MEASURE :
- No
- Pt
Choice of a temperature
measurement with or without
Pt100/Pt1000.
TYPE
:
- Auto.
- Manual
Choice between an automatic or a
manual temperature
measurement.
TEMP.
:
XX °C
Possibility to enter the sample
temperature in a manual
compensation.
COMP.
:
-
NERNST
Choice of the temperature
compensation type :
- linear compensation
(0.1984 mV/°C).
- Pure
- compensation according to the
ultrapure water curve.
- Matrix 1 - compensation according to the
sulphate curve (4.84 mg/l
corresponds to a pH 4.0 to 25°C).
- Matrix 2 - compensation according to the
ammoniac/hydrazine curve (0.272
mg/l am 20 µg/l
corresponds to a pH 9.0 to 25°C).
- Matrix 3 - compensation according to the
ammoniac / morpholine /
hydrazine curve (1.832 mg/l
am 10 mg/l morp
50 µg/l hydrazine corresponds to
a pH 9.6 to 25°C).
- Matrix 4 - compensation according to the
phosphate curve (3 mg/l
phos 0.3 mg /l
ammoniac).
Coeff.
- adjustable coefficient (pH/10°C)
∆
:
XX
In the case of a programmable
coefficient, possibility to enter the
value of the coefficient (value in
pH/10°C or pH/18°F).
MEASURE :
TYPE : Manual
TEMP. : 20°C
COMP. : Coeff.
∆
. : 00.010 pH
TEMP. COMP.
Select
Yes