Transmitter 9135 – pH/redox measurement
Page 37
Calibration
This menu is dedicated to the calibration of the
electrode used.
Press ENTER to validate the choice.
Redox calibration
Follow the pH calibration procedure in § 4.
Note :
In redox calibration the measure unit is mV.
LIMITS :
- zero shift :
±
250 mV
- slope shift : 70 … 120 %
TEMP CALIB.
PARAMETERS
CALIBRATION
Select
REDOX CALIB.
PROGRAMMING
REDOX CALIB.
Select
EXECUTION