Hach PHC108 Скачать руководство пользователя страница 11

5. 

 Shake the probe

from side to side to
refresh the reference
junction and remove
air bubbles.

6. 

 Stir gently, then

read the pH value of
the buffer solution.
The meter accepts
or rejects the result.

Section 8 Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time and life of the probe.
Keep the probe in the recommended storage solution between measurements.

8.1 Clean the probe

Clean the probe regularly to remove contamination and to keep the reference junction open.
Symptoms of contamination:

• Incorrect or irregular readings
• Slow stabilization times
• Calibration errors
• Sample material stays on the probe

1.

Rinse the probe with deionized water. Use warm (35–45 °C (95–113 °F)) deionized water to
remove storage solution that dries on the probe. Dry the probe body with a lint-free cloth.

2.

Soak the glass bulb and reference junction in the applicable cleaning solution for the specified
time. Refer to 

Table 1

 and 

Consumables

 on page 11.

3.

Rinse or soak the probe for 1 minute in deionized water. Dry the probe body with a lint-free cloth.

4.

Soak the probe in pH 4 buffer for 20 minutes.

5.

Rinse the probe with deionized water. Dry the probe body with a lint-free cloth.

Table 1  Cleaning solution

Contamination

Cleaning solution

Active component

Soak time

General
contamination

Electrode cleaning solution
for regular maintenance

KATHON

 CG,

DECONEX

®

11 

12–16 hours

Minerals

Electrode cleaning solution
for minerals/inorganic
contamination

Phosphoric acid
(~10%)

10–15 minutes

Fats, grease and oils

Electrode cleaning solution
for fats, oils and grease
contamination

KATHON

 CG,

TRITON

®

 X

2 hours maximum

Proteins

Electrode cleaning solution
for proteins/organic
contamination

Pepsin in HCl

3 hours maximum

English

   

9

Содержание PHC108

Страница 1: ...DOC022 53 80428 PHC108 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample requirements 7 6 2 Sample measurement notes 7 6 3 Sample measurement procedure 8 Section 7 Verify...

Страница 4: ...Table of Contents 2...

Страница 5: ...e overview 1 Probe soaker bottle with storage solution 3 Reference junction 2 Cable 4 Glass sensor and temperature sensor Section 2 Specifications Specifications are subject to change without notice S...

Страница 6: ...CC ISED Section 3 Safety information 3 1 Intended use The Intellical PHC108 probe is intended for use by individuals who measure the pH of semi solid samples emulsions and slurries in the laboratory T...

Страница 7: ...ith a soaker bottle that contains storage solution to keep the glass sensor and reference junctions hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...on the sensing element Make sure that the probe stays clean if proteins are in the sample Do not measure samples that are hotter or colder than the specified operating temperature of the probe Refer t...

Страница 10: ...ffer solution to make sure the result is accurate The meter compares the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solu...

Страница 11: ...robe Dry the probe body with a lint free cloth 2 Soak the glass bulb and reference junction in the applicable cleaning solution for the specified time Refer to Table 1 and Consumables on page 11 3 Rin...

Страница 12: ...ch solution 2 Rinse the probe with deionized water Blot dry with a lint free cloth 3 Calibrate the probe 8 3 Storage N O T I C E Probes can become permanently damaged if kept in a storage solution tha...

Страница 13: ...les Increase the sample temperature or adjust the temperature of different samples to be the same within 2 C 3 6 F Procedure problem causes slow stabilization and prevents accurate calibrations or mea...

Страница 14: ...solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769920 pH 7 00 and 10 01 buffer solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769820 pH color coded buffer solution kit NIST 5...

Страница 15: ...SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder 30...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: