Section 9 Troubleshooting

Keep the probe clean and in the recommended storage solution when not in use for the best
accuracy, stabilization time and life of the probe.

Problem

Possible cause

Solution

Decreased probe performance
causes slow stabilization and
prevents accurate calibrations
or measurements.

The glass sensor is dirty.

Clean and condition the probe.
Refer to 

Clean the probe

on page 9.

The glass sensor has become
dry.

Clean and condition the probe.
Refer to 

Maintenance

 on page 9.

The calibration slope of the
probe has changed.

Increase the accepted slope limit
settings if possible, or contact
technical support.

Sample properties cause slow
stabilization or inaccurate
measurements.

The sample absorbs carbon
dioxide (CO

2

) from the air,

which causes the pH value to
slowly decrease in low ionic
strength (LIS) or high purity
samples.

Use the LIS chamber for
LIS/high purity samples to
prevent CO

2

 absorption.

The sample temperature is
low, or there is a large
temperature difference
between samples.

Increase the sample temperature
or adjust the temperature of
different samples to be the same
(within 2 °C (3.6 °F)).

Procedure problem causes
slow stabilization and prevents
accurate calibrations or
measurements.

The probe is not conditioned
to the sample.

Soak the probe in the sample
before sample measurements.
Refer to 

Preparation for use

on page 5.

Air bubbles are around or
below the probe tip.

Carefully tap or shake the probe
to remove air bubbles.

The electrical connection
through the reference junction
is not sufficient.

Shake the probe in the solution
from side to side to refresh the
reference junction.

The stir speed is too slow or
too fast.

Try a different stir speed.

An incorrect buffer solution
was used or the buffer solution
has contamination.

Use the specified buffer solutions
of good quality.

Section 10 Consumables

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Description

Quantity

2

Item no.

Hach pH electrode storage solution

500 mL

2756549 

3 M KCl solution

50 mL

3

2756559 

2

Larger quantities are available for some items.

3

Larger quantities are available.

English

   

11

Содержание PHC108

Страница 1: ...DOC022 53 80428 PHC108 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample requirements 7 6 2 Sample measurement notes 7 6 3 Sample measurement procedure 8 Section 7 Verify...

Страница 4: ...Table of Contents 2...

Страница 5: ...e overview 1 Probe soaker bottle with storage solution 3 Reference junction 2 Cable 4 Glass sensor and temperature sensor Section 2 Specifications Specifications are subject to change without notice S...

Страница 6: ...CC ISED Section 3 Safety information 3 1 Intended use The Intellical PHC108 probe is intended for use by individuals who measure the pH of semi solid samples emulsions and slurries in the laboratory T...

Страница 7: ...ith a soaker bottle that contains storage solution to keep the glass sensor and reference junctions hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...on the sensing element Make sure that the probe stays clean if proteins are in the sample Do not measure samples that are hotter or colder than the specified operating temperature of the probe Refer t...

Страница 10: ...ffer solution to make sure the result is accurate The meter compares the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solu...

Страница 11: ...robe Dry the probe body with a lint free cloth 2 Soak the glass bulb and reference junction in the applicable cleaning solution for the specified time Refer to Table 1 and Consumables on page 11 3 Rin...

Страница 12: ...ch solution 2 Rinse the probe with deionized water Blot dry with a lint free cloth 3 Calibrate the probe 8 3 Storage N O T I C E Probes can become permanently damaged if kept in a storage solution tha...

Страница 13: ...les Increase the sample temperature or adjust the temperature of different samples to be the same within 2 C 3 6 F Procedure problem causes slow stabilization and prevents accurate calibrations or mea...

Страница 14: ...solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769920 pH 7 00 and 10 01 buffer solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769820 pH color coded buffer solution kit NIST 5...

Страница 15: ...SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder 30...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: