Section 6 Sample measurement

The procedure that follows is applicable to meters that can connect to Intellical pH probes. Refer to
the applicable meter documentation for meter operation and probe-specific settings.

6.1 Sample requirements

N O T I C E

 

Make sure that the sample type is compatible with the probe, or probe damage can occur.

• Do not use force to push the probe into samples. Use a stainless steel rod to make a hole in the

sample, then put the probe into the hole.

• Do not use the probe to measure samples where glass is not permitted.
• Probes with a Ag/AgCl reference system and single junction are not compatible with solutions that

contain silver complexing or binding agents (e.g., TRIS, proteins and sulfides). To measure in
these solutions, use a double junction probe, a probe with a Ag+ (silver ion) barrier or a Red Rod
probe.

• Proteins can collect on the sensing element. Make sure that the probe stays clean if proteins are in

the sample.

• Do not measure samples that are hotter or colder than the specified operating temperature of the

probe. Refer to 

Specifications

 on page 3.

6.2 Sample measurement notes

Read the notes that follow before sample measurements.

• Rinse the probe with deionized water and dry with a lint-free cloth between measurements to

prevent contamination.

• For faster stabilization, soak the probe for 3 or more minutes in the sample before measurements.
• If complete traceability is necessary, enter a sample ID and operator ID before measurement.

Refer to the meter manual for instructions.

• The meter automatically saves the measurement data when the user manually reads each data

point and when the meter is set to read at regular intervals. The user must manually save each
data point when the meter is set to read continuously.

• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure

to remove air bubbles before and during measurements.

English

   

7

Содержание PHC108

Страница 1: ...DOC022 53 80428 PHC108 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample requirements 7 6 2 Sample measurement notes 7 6 3 Sample measurement procedure 8 Section 7 Verify...

Страница 4: ...Table of Contents 2...

Страница 5: ...e overview 1 Probe soaker bottle with storage solution 3 Reference junction 2 Cable 4 Glass sensor and temperature sensor Section 2 Specifications Specifications are subject to change without notice S...

Страница 6: ...CC ISED Section 3 Safety information 3 1 Intended use The Intellical PHC108 probe is intended for use by individuals who measure the pH of semi solid samples emulsions and slurries in the laboratory T...

Страница 7: ...ith a soaker bottle that contains storage solution to keep the glass sensor and reference junctions hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...on the sensing element Make sure that the probe stays clean if proteins are in the sample Do not measure samples that are hotter or colder than the specified operating temperature of the probe Refer t...

Страница 10: ...ffer solution to make sure the result is accurate The meter compares the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solu...

Страница 11: ...robe Dry the probe body with a lint free cloth 2 Soak the glass bulb and reference junction in the applicable cleaning solution for the specified time Refer to Table 1 and Consumables on page 11 3 Rin...

Страница 12: ...ch solution 2 Rinse the probe with deionized water Blot dry with a lint free cloth 3 Calibrate the probe 8 3 Storage N O T I C E Probes can become permanently damaged if kept in a storage solution tha...

Страница 13: ...les Increase the sample temperature or adjust the temperature of different samples to be the same within 2 C 3 6 F Procedure problem causes slow stabilization and prevents accurate calibrations or mea...

Страница 14: ...solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769920 pH 7 00 and 10 01 buffer solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769820 pH color coded buffer solution kit NIST 5...

Страница 15: ...SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder 30...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: