• Use two or three buffer solutions for best results. Two buffer solutions are sufficient if the expected

sample pH is between the pH of the two buffer solutions. The sequence in which the pH buffer
solutions are used is not important. Use buffer solutions that are 2 or more pH units apart.

• For a one-point calibration, use a pH buffer near the expected sample pH.
• Use the default calibration options or change the options in the probe settings menu.
• Use the single display mode for calibration when more than one probe is connected to the meter (if

applicable).

• Calibrate the probes and verify the calibration regularly for best results. Use the meter to set

calibration reminders.

• The calibration data is stored in the probe. When a calibrated probe is connected to a different

meter with the same calibration options, a new calibration is not necessary.

• Air bubbles below the sensor when in solution can cause a slow response or error in the

calibration. Make sure to remove air bubbles during calibration.

• The pH buffer solutions have known pH values at different temperatures. The meter uses the mV

and temperature readings of the probe in the pH buffer solutions to calculate a calibration slope.
During measurements, the meter adjusts the slope for the sample temperature to determine the
pH value of the sample.

5.2 Calibration procedure

1. 

 Go to the

calibrate menu.
Select the probe, if
applicable. The
display shows the
pH buffer solutions
to use for calibration.

2. 

 Prepare or pour

the pH buffer
solutions in different
beakers.

3. 

 Rinse the probe

with deionized water.
Dry the probe with a
lint-free cloth.

4. 

 Put the probe in

the first pH buffer
solution. Make sure
that the sensor and
reference junction
are fully in the
solution. Do not put
the probe on the
bottom or sides of
the beaker.

5. 

 Shake the probe

from side to side to
refresh the reference
junction and remove
air bubbles.

6. 

 Stir slowly, then

read the pH value of
the buffer solution.
The display shows
the temperature-
corrected pH value
when the reading is
stable.

7. 

 Continue with

steps 

3

 through 

6

 for

the remaining
buffers or select
Done.

8. 

 Save the

calibration.

6

   

English

Содержание PHC108

Страница 1: ...DOC022 53 80428 PHC108 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample requirements 7 6 2 Sample measurement notes 7 6 3 Sample measurement procedure 8 Section 7 Verify...

Страница 4: ...Table of Contents 2...

Страница 5: ...e overview 1 Probe soaker bottle with storage solution 3 Reference junction 2 Cable 4 Glass sensor and temperature sensor Section 2 Specifications Specifications are subject to change without notice S...

Страница 6: ...CC ISED Section 3 Safety information 3 1 Intended use The Intellical PHC108 probe is intended for use by individuals who measure the pH of semi solid samples emulsions and slurries in the laboratory T...

Страница 7: ...ith a soaker bottle that contains storage solution to keep the glass sensor and reference junctions hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...on the sensing element Make sure that the probe stays clean if proteins are in the sample Do not measure samples that are hotter or colder than the specified operating temperature of the probe Refer t...

Страница 10: ...ffer solution to make sure the result is accurate The meter compares the selected pH buffer value to the measured pH value and accepts or rejects the measurement The user can change the pH buffer solu...

Страница 11: ...robe Dry the probe body with a lint free cloth 2 Soak the glass bulb and reference junction in the applicable cleaning solution for the specified time Refer to Table 1 and Consumables on page 11 3 Rin...

Страница 12: ...ch solution 2 Rinse the probe with deionized water Blot dry with a lint free cloth 3 Calibrate the probe 8 3 Storage N O T I C E Probes can become permanently damaged if kept in a storage solution tha...

Страница 13: ...les Increase the sample temperature or adjust the temperature of different samples to be the same within 2 C 3 6 F Procedure problem causes slow stabilization and prevents accurate calibrations or mea...

Страница 14: ...solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769920 pH 7 00 and 10 01 buffer solution kit Singlet one use packets 20 mL each 2 x 10 pkg 2769820 pH color coded buffer solution kit NIST 5...

Страница 15: ...SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker 30 mL plastic green 80 pkg SM5014 Beaker dispenser and holder 30...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: