36
Maintenance and Troubleshooting
Note:
During this step, it is very important to ensure
your finger does
not
come into contact with the
cathode (golden surface) as it could leave greasy
deposits on the surface.
33.
Take a silicone disc from the recharge kit,
hold it between thumb and forefinger and
position it on top of the anode.
Note:
As in the previous step, it is very important to
ensure your finger does
not
come into contact with the
cathode (golden surface) as it could leave greasy
deposits on the surface.
34.
Take a new cotton washer from the recharge
kit. Hold it between thumb and forefinger and
position it on top of the silicone disc.
Note:
The recharge cartridge container will be colored
black for all applications except for the 6110 TPA where
it will be colored white.
35.
Place the recharge cartridge container on a
flat work surface and, keeping the container
upright to avoid spilling any of the electrolyte
inside, carefully unscrew the top.
36.
Remove the packing component from the
center of the cartridge, and make sure that
the O-ring remains in place on top of the
cartridge. If it comes away then replace it
before continuing.
37.
If there are any visible bubbles in the
electrolyte, remove them using a stirring
motion with the packing component.
Содержание ORBISPHERE A1100
Страница 1: ...DOC024 52 93018 ORBISPHERE Model A1100 Oxygen Sensor USER MANUAL 10 2018 Edition 8...
Страница 2: ......
Страница 4: ...2 Table of Contents...
Страница 46: ...44 Accessories and Spare Parts...