14
Introduction
3.1.3 Sensor recharge kit
A recharge kit (as illustrated in
Figure 4
below) should have been ordered with the sensor as
this will be required to initially make the sensor operational. It is also required for sensor
cleaning and membrane replacement procedures.
Note:
The recharge kit for
oxygen
has a
blue
sticker on the front of the box.
The kit contains:
•
four recharge cartridges with pre-mounted membrane and electrolyte. The type of
membrane mounted in the cartridge will be specific to the kit ordered
•
two anode cleaning tools
•
a set of five cotton washers
•
a set of five silicone discs (available in all 2956A-XXX recharge kits only)
•
a set of replacement O-rings
•
a set of Dacron
®
mesh patches.
The black anode cleaning tool (
Figure 5
) is used to clean the anode of any deposits or residue
that may have formed. It is doubled-ended so it can be used for two membrane replacement
processes, each end being used once.
Figure 4 Sensor recharge kit
Figure 5 Anode cleaning tool