![Hach ORBISPHERE 6110 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/hach/orbisphere-6110/orbisphere-6110_user-manual_577089014.webp)
Option
Description
Clear audit trail
file
This option deletes the audit file.
Clear result file
This option deletes the package measurement data.
User management
1.
The list of registered users (up to a maximum of 99) for this instrument are displayed. To have the
Delete
and
Edit
options available, select one of the existing users.
Option
Description
New
To add a new user, all of the following fields must be entered, then select
Ok
to confirm:
•
Name
- Enter the user surname (3-15 characters)
•
First Name
- Enter the user first name (3-15 characters)
•
ID
- Enter an alphanumeric ID (1-10 characters)
•
Password
- Enter an alphanumeric password (3-15 characters)
•
Security level
- Select the security level from the drop-down list (see also the table below)
Edit or
Delete
To remove or modify an existing user, select the user in the user management screen and
select
Edit
or
Delete
as appropriate. To save all changes, press
Ok
or press the
Exit
button to
exit the screen without making any changes.
Level
Available options
Operator
Measurement and Analysis
Supervisor
Calibration, Measurement, Analysis and Maintenance
Manager
Configuration, Calibration, Measurement, Analysis and Maintenance
Administrator
Configuration, Calibration, Measurement, Analysis and Maintenance
Note: If access rights have been disabled (see
on page 13), all users are logged on at
Administrator level and there will be no name recorded against any actions in the audit trail.
Audit trail
View the list of user actions performed on the instrument. The actions are listed chronologically with
the latest action always at the top of the list. When the maximum of 1,000 recorded actions is
reached, the oldest is deleted and replaced with the newest.
Scroll through the audit trail screens using the
First
,
Previous
,
Next
and
Last
buttons, or select the
Exit
button to leave this option.
Note: To clear the audit trail file, use the option available in
on page 13.
Communication
This option sets the parameters required for exporting instrument data files to your PC or USB
storage device. Select from:
• Data download configuration (see
• Ethernet configuration (see
on page 15)
Data download configuration
Scroll through the list of available data files using the up/down arrow keys on the right, and select if
the file to be exported is in data or text format. By default, the text format is set to
YES
and the data
format to
NO
. Set both formats to
NO
if you do not require the file downloaded. Use the
Invert
Selection
button under each column to toggle between
YES
and
NO
. When the data formats have
been configured, select
Ok
to confirm.