Specification
Details
System alarm relay on the
main board
One system alarm relay; 1A-30 VAC or 0.5A-50 VDC on a resistance load.
Normally closed [NC] (NO relay also available) when instrument is turned on.
W A R N I N G
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS
Thermal cut off
Prevents ageing of sensors when exposed to high temperatures
Options
USB host; Ethernet 10/100 Base-T
Wall and pipe mount
instrument
(H x D x W)
236.5 x 160 x 250 mm; Weight 4.25 kg
9.31 x 6.30 x 9.84 in.; Weight 8.82 lbs
Panel mount instrument
(housing) (H x D x W)
156 (123) x 250 x 220 (214) mm; Weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in.; Weight 6.62 lbs
K-M1100 sensor
Specification
Details
Sample temperature Measurement from –5 to 50°C (23 to 122°F)
Sensor resistant to temperature from –5 to 100°C (23 to 212°F)
Sample pressure
1 to 20 bar absolute (14.5 to 290 psia)
Sample flow rate
requirement
50 to 300 mL/min
Sample types
K1100 sensor: Water only
M1100 sensor (low range): Water and beer
M1100 sensor (high range): Water, beer, wine, wort and carbonated drinks
For oxygen measurement
in
liquid phase
gas phase
Measurement range
Low range sensors
0 to 2000 ppb (dissolved.
Indicative values until
5000 ppb)
0 to 50 mbar or 0 to 5% O
2
(at
atm pressure)
High range sensors
0 to 40 ppm (dissolved)
0 to 1 bar or 0 to 100% O
2
(at
atm pressure)
Repeatability
Low range sensors
± 0.4 ppb or 1%,
whichever is the greater
± 0.01 mbar or 10 ppm gas or
1%, whichever is the greater
High range sensors
± 0.015 ppm or ± 2%,
whichever is the greater
± 0.4 mbar or 400 ppm gas or
2%, whichever is the greater
Reproducibility
Low range sensors
± 0.8 ppb or 2%,
whichever is the greater
± 0.02 mbar or 20 ppm gas or
2%, whichever is the greater
High range sensors
± 0.02 ppm or ± 3%,
whichever is the greater
± 0.5 mbar or 500 ppm gas or
3%, whichever is the greater
Accuracy
Low range sensors
± 0.8 ppb or 2%,
whichever is the greater
± 0.02 mbar or 20 ppm gas (at
atm pressure) or 2% of
reading, whichever is the
greater
High range sensors
± 0.02 ppm or ± 3%,
whichever is the greater
± 0.5 mbar or 500 ppm gas (at
atm pressure) or 3% of
reading, whichever is the
greater
4
English
Содержание Orbisphere 51 Series