![Hach Orbisphere 51 Series Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/hach/orbisphere-51-series/orbisphere-51-series_basic-user-manual_577087019.webp)
Option
Description
Display resolution
A maximum of 5 digits can be displayed. Decimals can be limited to 0, 1, 2 or 3 for
easier reading. The resolution affects only the data displayed, not the resolution of data
measured and stored.
T cut off
If this temperature is exceeded the measurement session is suspended and the system
displays a
HOT
alarm message. The system resumes when the temperature drops to
90% of the specified temperature. It is recommended to
Enable
this feature to maximize
sensor life and system performance.
T cut off value
Set to 5°C above the sample temperature.
K-M1100 advanced configuration
Note: The offset feature described below should be used for minor measurement adjustments only, and not as an
alternative to a sensor calibration. Make sure your sensor has been correctly calibrated before applying this feature.
1.
Select the
Advanced
button on the
Measurement configuration
screen:
Option
Description
Offset enabled
Check this box to enable the user measurement offset option. If checked enter an
offset value or a target value:
Offset value
Enter an offset value to manually adjust the measurement value. If the gas unit type
or gas unit (defined in the
Measurement configuration
screen) are changed the
offset value is automatically reset to zero.
Measurement
This field cannot be updated. It shows the current measurement value with the offset
applied.
Target value
Enter a target measurement value. The offset value is automatically computed so the
displayed measurement value will be equal to the target value.
Compute offset
Select this button to recompute the offset value at any time during the measurement
process. The offset value will be computed based on the current and target
measurement values.
Out of range
protection
Check this box to enable the out of range protection (recommended). When enabled
and the measured value exceeds the instrument specification, the measurement
interval will be increased to 1 minute to protect the lifetime of the sensor spot. If
disabled, the lifetime of the spot can be negatively impacted if the sensor is exposed
to high oxygen concentrations for long periods of time.
Measurement
interval
Set the value to between 2 and 60 seconds to define the interval for refreshing the
measurement value on the display.
Hold recovery time
This parameter defines the interval during which the outputs remain frozen after the
measurement is no longer on HOLD. Set the value to between OFF and 10 minutes,
according to the timing of your setup.
TC sensor configuration
1.
Select
Configure channel
from the
Measurement
menu:
Option
Description
Membrane
Sensor’s membrane number selection.
Medium
Liquid or gas phase.
Gas unit type
Partial, Fraction, Dissolved.
Gas unit
The list of available units depends on unit type selected above.
Note: This is the gas concentration measured by the sensor. When a composite unit is
selected (e.g. ppm » ppb) the unit will change depending on the range of the value to
display.
English
19
Содержание Orbisphere 51 Series
Страница 2: ...English 3 Deutsch 29 Français 58 Español 87 Čeština 115 Magyar 143 български 171 2 ...
Страница 201: ......