Connectors assembly instructions
D A N G E R
Electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the enclosure, use
only conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route cables in to the
instrument.
Cable gland wiring instructions
A waterproof cable gland is provided each time a cable must be connected inside the instrument.
The nickel-plated brass cable glands are EMC-types, designed so that the cable shields attach
directly to the instrument housing as a ground. Typical cable wiring instructions are detailed below.
1.
Unscrew the cable gland nut. Inside, the assembly is composed of a rubber gasket, and two
metal washers. Note that the ethernet gland on panel and wall mount instruments does not have
washers and the gasket is cut.
2.
If wiring a sensor cable, the cable has already been prepared so simply remove the piece of
plastic protection from the exposed shielding. For other cables, strip off external insulation as
required, and 25 mm of shielding. Strip the wires about 8 mm from their ends.
3.
Pass the cable through the nut, the rubber gasket, and the two washers.
4.
Pinch the shield so that its entire circumference is pressed between the two washers and pass
the cable into the housing, blocking the cable gland.
1
Cable
4
Wire
7
Gasket
2
Shield
5
O-ring
8
Gland nut
3
Instrument
6
Washers
N O T I C E
It is vitally important to ensure the shielding is pinched and secured between the two washers to ensure the
shielding attaches directly to the instrument housing as a ground. Failure to do this could cause damage to
the instrument, and for sensor cables give incorrect readings.
5.
Reattach and tighten the cable gland nut.
6.
Attach the wires to the corresponding terminal block connections.
Connection to mains power supply
Power supply connection (low voltage instruments)
For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 8-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
8
English
Содержание ORBISPHERE 410
Страница 179: ...ESD 2000 m 6562 ft 2000 m 179...
Страница 180: ...GFCI GFI PE 100 240 VAC 5 VDC PE PE 1 U 2 3 2 4 1 U 2 3 4 2 5 180...
Страница 181: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 1 1 5 USB A 2 USB B 4 6 2 3 Ethernet 7 1 4 8 181...
Страница 182: ...NEMA IP NEMA 4X IP66 1 Ethernet 2 25 mm 8 mm 3 4 1 4 7 2 5 8 3 6 5 6 10 30 VDC 8 BINDER 182...
Страница 184: ...1 4 2 3 1 2 3 3 1 2 3 4 1 2 3 4 3 4 5 4 2 6 1 7 ORBISPHERE J8 P8 4 185 J7 J8 6 186 7 186 184...
Страница 186: ...6 EC 7 TC 8 J7 9 J8 J7 16 J7 8 1 2 1 3 2 4 3 5 6 1 7 2 8 3 9 EC 9 TC J7 9 J7 12 10 11 12 13 16 J8 10 J8 9 EC TC 186...
Страница 188: ...32605 ORBISPHERE TC LED 10 1 menu Up Main Close Help Enter 188...
Страница 189: ...4 1 Configuration Security ID Clear ID ID ID 0 1 2 3 4 Access table Security 99 ID Add ID 4 1 4 189...
Страница 190: ...1 Configure View Conf Channel temperature Grid y z Auto scale Clean 2 190...
Страница 191: ...1 Config instrument Measurement Continuous 1 EC EC ppm ppb 5 0 1 2 3 HOT 90 2 TC TC EC TC 191...
Страница 195: ...Store once RAM TPO TPA Open data TPO data TPO TPA data TPA TPO TPA Hach 1 EC 2 EC TC EC 1 Modify O2 O3 195...
Страница 196: ...2 OK ideal current 3 O2 O2 30 O2 O2 O2 O3 196...
Страница 197: ...O3 O2 O2 O3 O2 O2 O3 O3 O3 TC 1 Modify Modify 2 OK ideal current 3 197...
Страница 198: ...ppm Cancel Validation Enable Enable Service done RS485 PROFIBUS DP USB HTTP TCP IP 198...
Страница 199: ...Hach 199...
Страница 224: ...2000 6562 2000 224...
Страница 225: ...GFCI GFI PE 100 240 5 1 U 2 3 4 225...
Страница 226: ...1 U 2 3 4 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 226...
Страница 227: ...1 1 5 USB A 2 4 USB B 6 2 3 Ethernet 7 1 4 8 NEMA IP NEMA 4X IP66 1 Ethernet 2 25 8 3 4 227...
Страница 228: ...1 4 7 2 5 8 3 6 5 6 10 30 8 BINDER 2 BINDER 1 10 30 2 3 4 5 6 10 30 7 10 30 8 3 228...
Страница 231: ...Ethernet RJ 45 ethernet ethernet 1 227 P3 4 6 7 J8 6 7 8 J7 9 J8 J7 16 J7 8 231...
Страница 233: ...J4 1 J5 2 J6 3 40 ORBISPHERE 29089 ORBISPHERE 32605 ORBISPHERE 233...
Страница 234: ...10 Menu Up Main Close Help CAP Enter Security 234...
Страница 244: ...last calibration variation 3 ppm Cancel Validation Services Enable Enable Service done 244...
Страница 245: ...Inputs Outputs RS485 PROFIBUS DP USB HTTP TCP IP Communications Products Global Configuration Hach 245...
Страница 356: ......
Страница 357: ......