Problem
Possible cause
Solution
The instrument
readings are not stable.
The analyzer was started
less than 2 hours ago.
Let the analyzer operate for 2 hours to become stable.
The electrode cable(s) are
cross-threaded on the
electrodes.
Disconnect the cable from the electrode. Push the
connector fully down on the electrode connector, then turn
the connector until finger tight.
The electrode cables are
connected to the wrong
electrodes.
Connect the blue cable to the reference electrode.
Connect the black cable to the sodium electrode.
The KCl electrolyte flow
rate is too fast.
Remove the reference electrode from the measurement
cell.
Push the collar down and turn the collar less than 1/4 of a
turn to lock the collar. When the collar is locked, the collar
does not turn. If the collar is not locked, the KCl electrolyte
will flow too fast from the reference electrode into the
measurement cell.
The measurements are
not accurate.
The analyzer is not
calibrated.
Calibrate the analyzer. Refer to
Do a calibration
in the
operations manual. For the best accuracy, calibrate the
analyzer at 7-day intervals (weekly).
Auto reactivation is
disabled or the interval
between reactivations is
more than 24 hours.
Start an auto reactivation. Push
diag
, then select
SENSOR REACTIVATION.
Note: For the best accuracy, set auto reactivation to on. Set the
interval to 24 hours. Refer to the operations manual.
The reactivation solution
bottle is empty.
Replace the reactivation solution. Refer to
calibration and reactivation solutions
on page 4.
There is a leak or
blockage in the
reactivation solution
tubing.
Look for a blockage or leak in the tubing. Replace the
tubing as necessary.
The reactivation solution
pump does not operate
correctly.
Push
diag
, then select PERFORM TEST > REACT SOL
DELIVERY > HIGH to set the reactivation solution pump
to a high flow rate. If reactivation solution does not flow to
the middle chamber of the measurement cell, replace the
reactivation solution pump.
The stir bar in the
overflow vessel does
not turn.
The mixer does not
operate correctly.
Note: The stir bar turns
intermittently.
Push
diag
, then select PERFORM TEST > MIXER to set
the mixer motor to on. If the stir bar does not move,
replace the mixer motor.
The KCl electrolyte
consumption is too
fast
1
.
The collar on the glass
part of the reference
electrode is not locked.
Remove the reference electrode.
Push the collar up. Push the collar down and turn the
collar less than 1/4 of a turn to lock the collar. Refer to the
illustrated steps in
Install the reference electrode
in the
installation manual.
When the collar is locked, the collar does not turn. If the
collar is not locked, the KCl electrolyte will flow too fast
from the reference electrode into the measurement cell.
1
The KCl electrolyte consumption is too fast if the level of the reservoir decreases by more than
3 mm (1/8 in.) in two days
20
English
Содержание NA5600 sc Na+
Страница 7: ...English 7...
Страница 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Страница 33: ...Deutsch 33...
Страница 38: ...38 Deutsch...
Страница 39: ...Entfernen der Referenzelektrode Ber cksichtigen Sie dabei die folgenden bebilderten Schritte Deutsch 39...
Страница 61: ...Italiano 61...
Страница 66: ...66 Italiano...
Страница 67: ...Rimozione dell elettrodo di riferimento Fare riferimento ai passaggi illustrati di seguito Italiano 67...
Страница 88: ...88 Fran ais...
Страница 93: ...Fran ais 93...
Страница 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Страница 115: ...Espa ol 115...
Страница 120: ...120 Espa ol...
Страница 121: ...Retirada del electrodo de referencia Consulte los pasos ilustrados que se muestran a continuaci n Espa ol 121...
Страница 142: ...142 Portugu s...
Страница 147: ...Portugu s 147...
Страница 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Страница 171: ...Nederlands 171...
Страница 176: ...176 Nederlands...
Страница 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Страница 198: ...198 Polski...
Страница 203: ...Polski 203...
Страница 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Страница 224: ...224 Svenska...
Страница 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Страница 250: ...250 Suomi...
Страница 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Страница 277: ...Magyar 277...
Страница 282: ...282 Magyar...
Страница 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Страница 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Страница 303: ...1 DIPA DIPA 303...
Страница 304: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 Merck Orion DIPA 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 304...
Страница 305: ...305...
Страница 307: ...4 KCl a b KCl 2 c 90 200 d 5 KCl a KCl 2 b c KCl 2 6 7 a menu REAGENTS STANDARDS b SET KCL LEVEL KCL 100 2 KCl 307...
Страница 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Страница 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Страница 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Страница 311: ...311...
Страница 312: ...312...
Страница 313: ...DIPA DIPA MSDS SDS 313...
Страница 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Страница 315: ...5 1 diag PERFORM TEST enter 2 Enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 315...
Страница 316: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 316...
Страница 319: ...CALIBRATION FAIL 301 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 319...
Страница 320: ...REACT VOLTAGE LOW 301 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 2 2 KCl 1 4 KCl 320...
Страница 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Страница 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Страница 335: ...T rk e 335...
Страница 340: ...340 T rk e...
Страница 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Страница 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Страница 361: ...1 DIPA DIPA 361...
Страница 362: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 L DIPA Merck Orion 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 DIPA 362...
Страница 363: ...363...
Страница 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Страница 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Страница 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Страница 369: ...369...
Страница 370: ...370...
Страница 371: ...DIPA DIPA MSDS SDS 371...
Страница 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Страница 373: ...5 1 diag PERFORM TEST enter 2 enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 373...
Страница 374: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 374...
Страница 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Страница 378: ...CALIBRATION FAIL 359 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 378...
Страница 379: ...REACT VOLTAGE LOW 359 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 379...
Страница 381: ...KCl 1 1 4 KCl KCl 2 1 4 KCl 381 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 1 KCl 3 mm 1 8 in 2 KCl 1 mm 381...
Страница 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Страница 389: ......