Problem
Possible cause
Solution
"UNSTABLE
READING" warning
occurs.
The rinsing time between
channel measurements is
too short.
Push menu, then select SETUP SYSTEM > CONFIGURE
ANALYZER > MEASUREMENT > SET MEASURE
CYCLE > MAX CYCLE TIME or CYCLE TIME. Increase
the MAX CYCLE TIME or CYCLE TIME setting to increase
the rinsing time.
"SAMPLE MISSING"
warning occurs.
The shut-off valve on the
sample line is closed.
Turn the shut-off valve to open.
The sample line is not
plumbed correctly.
Make sure that the sample line is connected to the correct
plumbing port. Refer to
Plumbing ports
in the installation
manual.
The sample flow valve is
closed.
Start the startup wizard. Push
menu
, then select SETUP
SYSTEM > STARTUP ANALYZER.
Push
enter
until prompted to adjust the sample flow rate.
Turn the sample flow valve counter-clockwise to increase
the flow rate to 6–9 L/hour on each channel.
The solenoid valve for the
channel does not operate
correctly.
Push
diag
, then select PERFORM TEST > SAMPLE
DRAIN to drain the overflow vessel.
Push
diag
, then select PERFORM TEST > SAMPLE
DELIVERY to open the sample valve. If sample does not
flow into the overflow vessel, examine the sample valve
wiring. Replace the sample valve if necessary.
"PH TOO LOW"
warning occurs.
The cover was removed
from the right chamber of
the measurement cell.
Install the cover on the right chamber of the measurement
cell.
The conditioning solution
bottle is less than 10% full
or empty.
Replace the conditioning solution. Refer to
There is an air leak in the
conditioning bottle cap.
• Make sure that the conditioning bottle cap is on tight.
• Examine the washer in the bottle cap. If wear or
damage is seen, replace the washer.
• Examine the plastic flange on the rim of the bottle. If
wear or damage is seen, replace the conditioning
solution bottle.
There is an air leak in the
conditioning tubing.
Look for a blockage or air leak in the conditioning tubing.
Replace tubing as necessary.
Note: The conditioning tubing contains gas, not solution.
The conditioning valve
does not operate correctly.
Push
diag
, then select PERFORM TEST >
CONDITIONING VALVE to open the conditioning valve. If
conditioning gas does not flow into the right chamber of
the measurement cell (seen as bubbles), replace the
conditioning valve.
18
English
Содержание NA5600 sc Na+
Страница 7: ...English 7...
Страница 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Страница 33: ...Deutsch 33...
Страница 38: ...38 Deutsch...
Страница 39: ...Entfernen der Referenzelektrode Ber cksichtigen Sie dabei die folgenden bebilderten Schritte Deutsch 39...
Страница 61: ...Italiano 61...
Страница 66: ...66 Italiano...
Страница 67: ...Rimozione dell elettrodo di riferimento Fare riferimento ai passaggi illustrati di seguito Italiano 67...
Страница 88: ...88 Fran ais...
Страница 93: ...Fran ais 93...
Страница 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Страница 115: ...Espa ol 115...
Страница 120: ...120 Espa ol...
Страница 121: ...Retirada del electrodo de referencia Consulte los pasos ilustrados que se muestran a continuaci n Espa ol 121...
Страница 142: ...142 Portugu s...
Страница 147: ...Portugu s 147...
Страница 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Страница 171: ...Nederlands 171...
Страница 176: ...176 Nederlands...
Страница 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Страница 198: ...198 Polski...
Страница 203: ...Polski 203...
Страница 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Страница 224: ...224 Svenska...
Страница 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Страница 250: ...250 Suomi...
Страница 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Страница 277: ...Magyar 277...
Страница 282: ...282 Magyar...
Страница 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Страница 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Страница 303: ...1 DIPA DIPA 303...
Страница 304: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 Merck Orion DIPA 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 304...
Страница 305: ...305...
Страница 307: ...4 KCl a b KCl 2 c 90 200 d 5 KCl a KCl 2 b c KCl 2 6 7 a menu REAGENTS STANDARDS b SET KCL LEVEL KCL 100 2 KCl 307...
Страница 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Страница 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Страница 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Страница 311: ...311...
Страница 312: ...312...
Страница 313: ...DIPA DIPA MSDS SDS 313...
Страница 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Страница 315: ...5 1 diag PERFORM TEST enter 2 Enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 315...
Страница 316: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 316...
Страница 319: ...CALIBRATION FAIL 301 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 319...
Страница 320: ...REACT VOLTAGE LOW 301 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 2 2 KCl 1 4 KCl 320...
Страница 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Страница 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Страница 335: ...T rk e 335...
Страница 340: ...340 T rk e...
Страница 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Страница 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Страница 361: ...1 DIPA DIPA 361...
Страница 362: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 L DIPA Merck Orion 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 DIPA 362...
Страница 363: ...363...
Страница 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Страница 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Страница 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Страница 369: ...369...
Страница 370: ...370...
Страница 371: ...DIPA DIPA MSDS SDS 371...
Страница 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Страница 373: ...5 1 diag PERFORM TEST enter 2 enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 373...
Страница 374: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 374...
Страница 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Страница 378: ...CALIBRATION FAIL 359 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 378...
Страница 379: ...REACT VOLTAGE LOW 359 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 379...
Страница 381: ...KCl 1 1 4 KCl KCl 2 1 4 KCl 381 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 1 KCl 3 mm 1 8 in 2 KCl 1 mm 381...
Страница 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Страница 389: ......