Problem
Possible cause
Solution
"CALIBRATION FAIL"
warning occurs.
The value of the
calibration solution in the
auto calibration settings is
different from the value on
the calibration solution
bottle.
Change the calibration solution value in the calibration
menu to the value on the calibration solution bottle. Refer
to
Set the auto calibration settings
in the operations
manual.
The calibration solution
bottle is empty.
Replace the calibration solution. Refer to
calibration and reactivation solutions
on page 4.
There is a leak or
blockage in the calibration
solution tubing.
Look for a blockage or leak in the tubing. Replace the
tubing as necessary.
The calibration solution
pump does not operate
correctly.
Push
diag
, then select PERFORM TEST > SAMPLE
DRAIN to fully drain the overflow vessel.
Push
diag
, then select PERFORM TEST > CAL STD
DELIVERY to set the calibration solution pump to on. If
calibration solution does not flow to the overflow vessel,
replace the calibration solution pump.
The stir bar does not turn.
Note: The stir bar turns
intermittently.
Push
diag
, then select PERFORM TEST > MIXER to set
the mixer motor to on. If the stir bar does not move,
replace the mixer motor.
The stir bar is not in the
overflow vessel.
Make sure that the stir bar is in the overflow vessel.
"REACT VOLTAGE
LOW" warning occurs.
The reactivation solution
bottle is empty.
Make sure that the end of the reactivation solution tube is
at the bottom of the bottle.
Replace the reactivation solution. Refer to
calibration and reactivation solutions
on page 4.
The sodium electrode
does not operate correctly.
• Make sure that the sodium electrode is installed
correctly. Refer to the installation manual.
• Make sure that there is no air in the glass bulb of the
sodium electrode.
The reference electrode
does not operate correctly.
• Remove the reference electrode from the measurement
cell. Lift the collar that is on the glass part to let some
KCl electrolyte flow out. Push the collar down and turn
the collar less than 1/4 of a turn to lock the collar. When
the collar is locked, the collar does not turn. Install the
reference electrode in the measurement cell.
• Make sure that the reference electrode is installed
correctly. Refer to the installation manual.
• To identify if the problem is fixed, push
diag
, then
select SENSOR REACTIVATION to do a reactivation.
There is a leak or
blockage in the
reactivation solution
tubing.
Look for a blockage or leak in the tubing. Replace tubing
as necessary.
The reactivation solution
pump does not operate
correctly.
Push
diag
, then select PERFORM TEST > REACT SOL
DELIVERY > HIGH to set the reactivation solution pump
to a high flow rate. If reactivation solution does not flow to
the middle chamber of the measurement cell, replace the
reactivation solution pump.
English
19
Содержание NA5600 sc Na+
Страница 7: ...English 7...
Страница 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Страница 33: ...Deutsch 33...
Страница 38: ...38 Deutsch...
Страница 39: ...Entfernen der Referenzelektrode Ber cksichtigen Sie dabei die folgenden bebilderten Schritte Deutsch 39...
Страница 61: ...Italiano 61...
Страница 66: ...66 Italiano...
Страница 67: ...Rimozione dell elettrodo di riferimento Fare riferimento ai passaggi illustrati di seguito Italiano 67...
Страница 88: ...88 Fran ais...
Страница 93: ...Fran ais 93...
Страница 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Страница 115: ...Espa ol 115...
Страница 120: ...120 Espa ol...
Страница 121: ...Retirada del electrodo de referencia Consulte los pasos ilustrados que se muestran a continuaci n Espa ol 121...
Страница 142: ...142 Portugu s...
Страница 147: ...Portugu s 147...
Страница 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Страница 171: ...Nederlands 171...
Страница 176: ...176 Nederlands...
Страница 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Страница 198: ...198 Polski...
Страница 203: ...Polski 203...
Страница 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Страница 224: ...224 Svenska...
Страница 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Страница 250: ...250 Suomi...
Страница 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Страница 277: ...Magyar 277...
Страница 282: ...282 Magyar...
Страница 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Страница 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Страница 303: ...1 DIPA DIPA 303...
Страница 304: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 Merck Orion DIPA 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 304...
Страница 305: ...305...
Страница 307: ...4 KCl a b KCl 2 c 90 200 d 5 KCl a KCl 2 b c KCl 2 6 7 a menu REAGENTS STANDARDS b SET KCL LEVEL KCL 100 2 KCl 307...
Страница 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Страница 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Страница 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Страница 311: ...311...
Страница 312: ...312...
Страница 313: ...DIPA DIPA MSDS SDS 313...
Страница 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Страница 315: ...5 1 diag PERFORM TEST enter 2 Enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 315...
Страница 316: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 316...
Страница 319: ...CALIBRATION FAIL 301 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 319...
Страница 320: ...REACT VOLTAGE LOW 301 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 2 2 KCl 1 4 KCl 320...
Страница 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Страница 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Страница 335: ...T rk e 335...
Страница 340: ...340 T rk e...
Страница 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Страница 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Страница 361: ...1 DIPA DIPA 361...
Страница 362: ...3 10 DIPA 28 MSDS SDS DIPA 99 1 L DIPA Merck Orion 1 MSDS SDS 2 menu STOP ANALYZER START ANALYZER 3 4 1 8 3 6 DIPA 362...
Страница 363: ...363...
Страница 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Страница 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Страница 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Страница 369: ...369...
Страница 370: ...370...
Страница 371: ...DIPA DIPA MSDS SDS 371...
Страница 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Страница 373: ...5 1 diag PERFORM TEST enter 2 enter YES 3 SAMPLE DRAIN enter 4 150 5 enter 6 back 7 home 1 6 2 3 4 5 6 373...
Страница 374: ...6 1 2 3 4 5 menu SETUP SYSTEM DISPLAY SETUP ADJUST ORDER SEE CURRENT ORDER back ADD MEASUREMENTS 374...
Страница 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Страница 378: ...CALIBRATION FAIL 359 diag PERFORM TEST SAMPLE DRAIN diag PERFORM TEST CAL STD DELIVERY diag PERFORM TEST MIXER 378...
Страница 379: ...REACT VOLTAGE LOW 359 KCl 1 4 diag SENSOR REACTIVATION diag PERFORM TEST REACT SOL DELIVERY HIGH 379...
Страница 381: ...KCl 1 1 4 KCl KCl 2 1 4 KCl 381 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 1 KCl 3 mm 1 8 in 2 KCl 1 mm 381...
Страница 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Страница 389: ......