Page 14
Reagent Consumption
6200018theory.fm
Section 2
4.
A portion of the solution in the mixer is aspirated out and the
buffer/indicator reagent is added. This solution aliquot and the reagent are
both dispensed back into the mixer.
5.
The solution in the mixer may now start to turn red.
6.
Some of the solution is aspirated from the mixer and sent to the detector
to prime the reaction coil.
7.
The rest of the solution is aspirated from the mixer, followed by a slug of
titrant reagent, and sent to the detector for titrimetric measurement using
a 612 nm LED and a 600 nm filter.
8.
As the titrant disperses in the reaction coil, a plug of blue solution may be
seen going through the detector.
The use of inflection points in the titrimetric measurement eliminates the need
for a reference measurement, even for colored samples.
Table 2
lists the port
numbers and their function.
2.2 Reagent Consumption
The rate of reagent use depends on many factors, including the number of
times the instrument cycles, calibration frequency, and the number of clean
cycles.
Table 3
provides the approximate volumes (in mL) of the reagents
consumed with each type of function per cycle.
Table 2 Port Functions For High Range Hardness Analysis
Port #
Function
Port #
Function
1
Detector
6
Waste
2
Sample 1
7
Reagent 3
3
Sample 2
8
Reagent 2 (indicator)
4
Standard 1
9
Reagent 1
5
Standard 2
10
Mixer
Table 3 Reagent Consumption in mL
Solution
Measurement
Calibration
Cleaning
Sample
5.07
0
2.0
Reagent 1
0.140
1.12
0
Reagent 2
0.16
2.56
0
Reagent 3
0.10
1.6
0
Standard 1
0
64.2
0
Standard 2
0
35.4
0
Cleaning Solution
0
0
6.8
Содержание Lange APA 6000
Страница 3: ...Page 3 6200018safety fm Safety Precautions Safety Precautions...
Страница 6: ...Visit http www hach com...
Страница 12: ...Visit http www hach com...
Страница 28: ...Visit http www hach com...