Theory of Operation

Section 2

Theory of Operation

The APA 6000™ High Range Hardness Process Analyzer is designed to 
provide reproducible results through consistent solution handling. Every step 
in an analysis is performed the same way each time, ensuring accurate 
results. Accuracy can be verified and established by measuring grab samples 
with a known concentration of analyte.

These steps describe how the analyzer works:

1.

The instrument aspirates sample into the holding coil.

2.

Appropriate reagents are aspirated.

3.

Sample and reagents are premixed in a mixing chamber.

4.

The sample-reagent mixture is dispensed into the reaction coil, then 
the detector.

5.

The mixture flows through the detector and the color is measured.

2.1 Method of Analysis

The sample is treated with a Magnesium-CDTA masking agent at acidic pH, 
to eliminate both carbonate/bicarbonate and heavy metal interferences. The 
treated sample is then mixed with an indicator/buffer solution at a higher pH.

Some of the sample/buffer/indicator mixture is aspirated from the mixer and 
delivered to the detector via the reaction coil. The remainder of the mixture 
then a slug of EDTA titrant is aspirated and delivered to the detector via the 
reaction coil.

As the mixture travels through the reaction coil, the titrant becomes dispersed 
through the sample/buffer/indicator mixture on both sides of the titrant slug, 
and forms a gradient of EDTA concentration. There will be a small volume of 
fluid on both sides of the titrant peak and where the EDTA quantitatively binds 
all hardness present, the indicator changes color.

As the fluid flows through the detector, its absorbance is recorded, producing 
a characteristic titration peak. The peak width is determined from the 
inflection point on both sides of the peak. The instrument calibrates using two 
standards to establish a multi-point calibration curve.

In the analysis, segments of untreated sample are kept acidic to assure that 
no precipitation occurs in the system.

The measurement cycle follows these basic steps:

1.

The analyzer rinses out the holding coil, mixer, and reaction coil in two 
steps with acidified sample.

2.

Sample is pulled into the holding coil, followed by masking reagent.

3.

This mixture is dispensed into the mixer and stirred. The mixer is filled 
in two strokes of the autoburette to assure only representative sample 
is in the mixer. 

Содержание Lange APA 6000

Страница 1: ...Catalog Number 62000 18 APA 6000 High Range Hardness Process Analyzer Operation Manual Hach Company 2002 2005 All rights reserved Printed in the U S A eac April 2005 Edition 2...

Страница 2: ...Tubing Lengths 8 1 2 7 Vortex Sample Conditioning Block 11 1 2 8 High Range Hardness Reagents 11 Section 2 Theory of Operation 13 2 1 Method of Analysis 13 2 2 Reagent Consumption 14 Section 3 Sequenc...

Страница 3: ...Page 3 6200018safety fm Safety Precautions Safety Precautions...

Страница 4: ...whichever is greater Repeatability 5 of reading or 2 mg L as CaCO3 whichever is greater Response Time For single channel instruments less than 17 minutes for 90 response to step change at sample inlet...

Страница 5: ...eligrosa Revise las fichas de seguridad de materiales y familiar cese con los procedimientos de seguridad antes de manipular productos qu micos GEFAHR Das Arbeiten mit chemischen Proben Standards und...

Страница 6: ...Visit http www hach com...

Страница 7: ...ace keypad and display The reagent containers are sized to provide approximately 30 days of continuous operation The reagents are factory filled in one liter containers and meet Hach quality standards...

Страница 8: ...rength drops until only sample indicator is present again The reaction coil is placed between the valve and detector After the dispersion described above occurs the response is measured as fluids pass...

Страница 9: ...ncia siempre estudie las Hojas de Datos de Seguridad de los Materiales antes de manipular recipientes dep sitos y sistemas de entrega que contengan reactivos y patrones qu micos Siempre se recomienda...

Страница 10: ...el tubing 0 03 ID x 0 062 OD 152 5 cm 60 in Port 7 Reagent 3 8 Tefzel tubing 0 03 ID x 0 062 OD 152 5 cm 60 in Port 8 Reagent 2 9 Tefzel tubing 0 03 ID x 0 062 OD 145 cm 57 in Port 9 Reagent 1 10 Tefz...

Страница 11: ...hat adjusts the pH of the sample Reagent 2 is an indicator used to determine the endpoints of the titration It is supplied as a kit containing a solution and a powder The components must be mixed just...

Страница 12: ...Visit http www hach com...

Страница 13: ...tor via the reaction coil The remainder of the mixture then a slug of EDTA titrant is aspirated and delivered to the detector via the reaction coil As the mixture travels through the reaction coil the...

Страница 14: ...of inflection points in the titrimetric measurement eliminates the need for a reference measurement even for colored samples Table 2 lists the port numbers and their function 2 2 Reagent Consumption T...

Страница 15: ...0 Aspirate Mixer 1000 Dispense Mixer 950 Aspirate Mixer 100 Aspirate Waste 1050 Dispense Sample 1500 Aspirate Mask acid 50 Aspirate Mixer 1000 Dispense Mixer 950 Aspirate Mixer 100 Aspirate Detector 6...

Страница 16: ...irate Detector 1600 Dispense Between ports 1600 Aspirate Mixer 1600 Dispense Between ports 400 Aspirate Mixer 400 Dispense Between ports 1600 Aspirate Detector 1600 Dispense Mixer 1600 Aspirate Detect...

Страница 17: ...e Between ports 1600 Aspirate Waste 1600 Dispense Mask acid 1000 Aspirate Waste 1000 Dispense Buffer Indicator 1000 Aspirate Waste 1000 Dispense Titrant 1000 Aspirate Waste 1000 Dispense Between ports...

Страница 18: ...er 100 Aspirate Detector 600 Dispense Waste 1000 Dispense Standard 1 1600 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 1 1495 Aspirate Standard 2 85 Aspirate Mixer 0 Depressurize Mask acid...

Страница 19: ...1600 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 1 1410 Aspirate Standard 2 170 Aspirate Mixer 0 Depressurize Mask acid 20 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 1 760 Asp...

Страница 20: ...Standard 2 1600 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 1 1580 Aspirate Mixer 0 Depressurize Mask acid 20 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 1 760 Aspirate Mask a...

Страница 21: ...Standard 2 1600 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 2 1580 Aspirate Mixer 0 Depressurize Mask acid 20 Aspirate Mixer 850 Dispense Waste 750 Dispense Standard 2 760 Aspirate Mask a...

Страница 22: ...nu press EXIT The calibration is cubic and is based on 50 100 500 and 1000 mg L as CaCO3 standards that are prepared by the analyzer from the 0 and 1000 mg L as CaCO3 standards Data displayed for the...

Страница 23: ...pipet to transfer 20 00 mL of Standard 2 solution into a clean 250 mL Erlenmeyer flask 3 Dilute to about the 100 mL mark with deionized water 4 Using a marked dropper add 1 mL of Reagent 1 Swirl to mi...

Страница 24: ...nd pipet size for the buret titration procedure 9 Repeat steps 1 8 twice 10 Use the average value for the titer of Reagent 3 Repeat steps 1 8 twice Average the three values for the titer of Reagent 3...

Страница 25: ...metric pipet transfer the sample volume from Table 10 into a clean 250 mL Erlenmeyer flask 6 Dilute to about the 100 mL mark with deionized water 7 Using a marked dropper add 1 mL of Reagent 1 Swirl t...

Страница 26: ...ionized water if necessary 5 Using a marked dropper add 1 mL of Reagent 1 Swirl to mix 6 Using a graduated cylinder add 8 mL of Reagent 2 Swirl to mix Note Prepare Reagent 2 by mixing one bottle of In...

Страница 27: ...t 25 mL Class A certified each 26365 40 Cylinder graduated 10 mL 1 each 508 38 Delivery tubes for the Digital Titrator 90 with J hook 5 pkg 41578 00 Digital Titrator each 16900 01 Digital Titrator car...

Страница 28: ...Visit http www hach com...

Отзывы: