English
13
Error messages and display messages
Maintenance and service
Change the air filter (every year or if it is very dirty):
•
Turn the instrument off and disconnect the power cord.
•
Remove the fitted square air filter from the bottom of the instrument
and replace the dirty filter mat with a new one.
Clean the device
The outside of the instrument can be cleaned with a dry cloth or a mild
household cleaning agent. Under no circumstances may solvents such
as petroleum spirit, acetone or similar substances be used.
1.
Turn the instrument off and disconnect the power cord.
2.
When the instrument is cool, clean the surface of the instrument
with a soft, moist cloth and a weak soap solution. Make sure that
no water gets into the instrument.
Error message
Definition
Resolution
Prog.Mem. full
The user-specific program
memory is full
Delete a digestion program
that is no longer required
Prog.Mem. Clear
The user-specific program
memory is empty
Enter a user-specific
digestion program
Name not
recognised
No name has been entered
Designation
Name already
used esc
A name has been assigned
twice
Select a different name
Close the lid
The lid has not been closed
Close the lid
ERR: 1000
temp. too high
Permissible temperature
exceeded
Contact the Service
department
ERR: 1001
Catch
Lock faulty
Contact the Service
department
N O T I C E
Check the condition of your air filter (level of dirt) on a regular basis, as a very
dirty filter mat lengthens the cooling-down phase of the thermostat.
D A N G E R
Potential danger with contact with chemical/biological substances.
Working with chemical samples, standards and reagents can be dangerous.
Make yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before the work and read and follow all relevant safety
data sheets.
WA R N I N G
Fire hazard! Do not use flammable agents to clean the instrument.
To prevent instrument damage, do not use cleaning agents such as turpentine,
acetone or similar products to clean the instrument or the display.
C A U T I O N
Burn hazard
Do not clean the instrument when it is hot.
Содержание HT 200 S
Страница 52: ...Français 49 ...
Страница 86: ...Български 22 ...
Страница 104: ...Čeština 40 ...
Страница 220: ...Magyar 156 ...
Страница 238: ...Italiano 174 ...
Страница 256: ...Lietuvių k 192 ...
Страница 274: ...Македонски 210 ...
Страница 292: ...Nederlands 228 ...
Страница 326: ...Português 262 ...
Страница 344: ...Română 280 ...
Страница 380: ...Slovensky 316 ...
Страница 398: ...Slovenščina 334 ...
Страница 447: ......