Digestion end
Standard programs 100 and HT, and user-specific
digestion operations
The cooling-down phase (automatic rapid cooling) begins immediately
after the digestion time has elapsed and after the digestion is aborted by
pressing
key 4
under
Esc
. As soon as a temperature of
approx. 80 °C
is reached, the lock is automatically released; the
Close flap
instruction
on the display must always be observed. The end of the cooling-down
phase or the abortion of the digestion is signaled by an acoustic sound
and by the
Digestion over!
message on the display.
Press
key 4
under
ESC
to return to the main menu.
Standard program COD
The cooling-down phase is stopped at a temperature of
approx. 100 °C
and the lid lock is released. This is indicated by an acoustic signal.
The
Please invert cuvette
message appears on the display.
After the cuvettes have been inverted and placed back in the
thermostats, and the lid has been closed,
key 3
under the
arrow
symbol
can be pressed to continue cooling to a temperature of approx.
44 °C. In this instance, the lid is locked again until the temperature has
cooled to 80 °C.
The end of the cooling-down phase or the abortion of the digestion is
indicated by an acoustic signal and by the
Digestion over!
message on
the display.
Press
key 4
under
Esc
to return to the main menu.
User programs
There are nine program locations available that can be configured by the
user.
Programming (PRG) of user-specific temperature programs
Figure 7 Programming
User-specific digestion programs are programmed using the
PRG
menu.
The
digestion term
,
temperature
,
digestion time
and
position
must
be defined or calculated by the user.
Enter the digestion term
Press
key 1
to set the saved alphanumeric character set to the required
number, letter or space.
Press
key 2
to move the character shown as
_
on the display one place
to the right or back to the starting position.
A maximum of
three
characters can be entered.
N O T I C E
Take care when inverting — the cuvettes are hot! Wear protective clothing,
protective gloves and safety glasses/face protection that are suitable for the work
being performed.
1
Temperature
menu
4
Actuate key to select
VAR
. User-
specific digestion programs are
varied using the VAR menu.
2
Press key to return to the
Main
menu
.
5
Press key to select
PRG
. User-
specific digestions can be
programmed using the PRG
menu.
3
Press key to select
CPRG
. User-
specific digestion programs can
be deleted using the CPRG
menu.
1
2
5
3
4
Содержание HT 200 S
Страница 52: ...Français 49 ...
Страница 86: ...Български 22 ...
Страница 104: ...Čeština 40 ...
Страница 220: ...Magyar 156 ...
Страница 238: ...Italiano 174 ...
Страница 256: ...Lietuvių k 192 ...
Страница 274: ...Македонски 210 ...
Страница 292: ...Nederlands 228 ...
Страница 326: ...Português 262 ...
Страница 344: ...Română 280 ...
Страница 380: ...Slovensky 316 ...
Страница 398: ...Slovenščina 334 ...
Страница 447: ......