Table 1 General specifications (continued)
Specification
Details
UKEX—GS2440EX
Certificate: ITS22UKEX0506X
II 1 G Ex ia IIC T4 Ga
-20 °C ≤ Ta ≤ +60 °C
IECEx—GS2440EX
Certificate: IECEx ETL 22.0003X
Ex ia IIC T4 Ga
Tamb: -20 to +60 °C
North America—GS2440EX
Certificate: ETL21CA104799609X
Class I Zone 0 AEx ia IIC T4 Ga
Class I Division 1 Groups A–D T4
Ex ia IIC T4 Ga
Tamb: -20 to +60 °C
Warranty
1 year
Table 2 Performance specifications
Specification
Details
Measurement range in water 0 to 5 mg/L H
2
S
Measurement range in air
0 to 1000 ppm H
2
S
Accuracy
± 5% full scale (discrete measurement accuracy, 90-second purge of
1000 ± 20 ppm H
2
S standard after one-point calibration)
Response time (t90)
< 30 seconds
Detection limit
1% full scale
Table 3 Interferences
Gas
Interference
level
in gas (%)
1
Interference
level
in liquid (%)
2
Gas
Interference
level
in gas (%)
1
Interference
level
in liquid
(%)
2
Methane (CH
4
)
0
0
Hydrogen (H
2
)
0.8
96
Carbon dioxide (CO
2
)
0
0
Carbon monoxide (CO)
0.6
77
Nitrogen (N
2
)
0
0
Dimethyl sulfide
((CH
3
)
2
S))
18
18
Oxygen (O
2
)
0
0
Methyl mercaptan
(CH
3
SH)
174
44
Air (O
2
, N
2
, Ar)
0
0
Ethyl mercaptan
(C
2
H
6
S)
13
14
Nitrous oxide (N
2
O)
0
0
Sulfur dioxide (SO
2
)
40
1
Ammonia (NH
3
)
0
0
1
Given as a signal for the interfering species in % of H
2
S signal at equal partial pressures.
2
Given as a signal for the interfering species in % of H
2
S signal at equal molar concentrations.