English
27
Troubleshooting
Error displayed
Cause
Resolution
Test execution errors
Barcode label not read
Barcode faulty
Reinsert the cell.
If the barcode is not
recognised, contact
technical support.
Please close the cover.
Close the cover.
Absorbance > 3.5!
The measured
absorbance exceeds 3.5
Dilute sample and
measure again
Fault
Barcode check
number?
Please update program
data!
Deviation to the stored
data
Update program data
It’s recommended to
execute a Full System
Check
Check of the air values
failed
Switch the instrument off
and then back on again.
If the system check is not
successful, contact
technical support.
Blank value correction
not possible!
Blank value correction
not possible with
LCW919.
Fault
Program not
accessible.
Please update program
data!
Barcode test not present
Update program data
Fault
Clean cuvette!
The cuvette is dirty or
there are undissolved
particles in the cuvette
Clean the cuvette; allow
the particles to settle
Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]
Test program stops when
the instrument is started
Check the lamp and
replace it if necessary.
Close lid.
Press
Start Again
.
Fault
Test program stopped!
Please remove the
cuvette
Close the lid.
Test program stops when
the instrument is started
Remove the cuvette/
sample cell from the cell
compartment.
Press
OK
.
Error
Selfcheck stopped.
Hardware error.
Error [x]
Electronic defect
Contact technical support
and indicate the error
number
Error
Too much ambient
light!
Move instrument into
shade
or close the lid
The instrument sensors
detects too much
ambient light.
Decrease ambient light.
(Avoid direct sunlight.)
Close lid.
No help function is
available for this
program.
Shelf life exceeded!
Use chemicals?
The analysis is possibly
erroneous. Use new
chemicals
No evaluation!
Error in the test database
/ user database
Check programming
Contact technical support
No barcode!
No barcode found
Reinsert the cell.
If the barcode is not
recognised, contact
technical support.
Error displayed
Cause
Resolution
Содержание DR 6000
Страница 2: ...English 3 Espa ol 31 Fran ais 59 89 117 Portugu s 143 171...
Страница 87: ...Fran ais 87...
Страница 88: ...Fran ais 88...
Страница 90: ...90 CAN CSA C22 2 No 61010 1 2 1 2 II 2000 m 6560...
Страница 92: ...92 DR 6000 190 1100 nm 320 1100 nm 190 360 nm DR 6000...
Страница 93: ...93 DR 6000 DR 6000 DR 6000 USB RFID 1 DR 6000 LINK2SC...
Страница 94: ...94 10 40 C 80 15 cm 112 1 1 USB A 5 2 6 3 7 RFID 4...
Страница 95: ...95 2 1 95 2 100 240 V 50 60 Hz 3 On 95 1 4 USB B 2 5 USB A 3 6...
Страница 96: ...96 1 2 1 cm 3 1 4 3 1 2 3 45 20 DR 6000 1 2 OK...
Страница 97: ...97 1 2 45 3 4 OK 2 112 1 2 1...
Страница 98: ...98 1 YZ OK USB US USB 116 1 ABC abc 123 CE 1 1...
Страница 99: ...99 2 HACH LANGE HACH HACH LANGE HACH LANGE HACH HACH HACH LANGE 3 Abs T 4 1 2...
Страница 100: ...100 5000 ID ID 1 2 ID ID PC 2...
Страница 101: ...101 3 4 ID ID 4 5 OK USB DataLog XML Extensible Markup Language CSV Comma Separated Value DL csv DL xml...
Страница 102: ...102 1 USB DR 6000 USB A DR 6000 2 3 4 Options PC 5 6 OK...
Страница 103: ...103 200 Stored Programs Stored Programs 1 2 OK 3 4 Web 1 2 3...
Страница 104: ...104 4 5 Dilution 6 100 1 1 13mm ID 10 1 1 1 96...
Страница 105: ...105 Barcode Programs Dilution 3 1 Single Wavelength 2 Options...
Страница 106: ...106 3 100 Trans Abs 190 1100 nm 105 Read 1 1 5 1 7 100 100 3...
Страница 107: ...107 4 Multi Wavelength Options 4 100 Trans Abs 190 1100 nm 105 Read 1 1 5 1 7 100 105 4...
Страница 108: ...108 T Abs Wavelength Scan Options 5 View Table Graph View Table USB A 5...
Страница 109: ...109 1 1 Wavelength Scan 2 y Read 1 1 5 1 7 100 100 5...
Страница 110: ...110 3 Zero Zeroing 4 5 Read Reading x Options...
Страница 111: ...111 1 Time Course 2 Options 6 Scale Units y USB A Read 1 1 5 1 7 100 6...
Страница 112: ...112 Barcode label not read Please close the cover 3 5 3 5 Fault Barcode check number LCW919 Fault Program not accessible...
Страница 118: ...118 CAN CSA C22 2 No 61010 1 2 1 2 II 2000m 6560ft...
Страница 119: ...119 RFID RFID DR 6000 RFID DR 6000 RFID DR 6000 RFID RFID 13 56MHz 2005 8 12 EU 2002 96 EC EU UV UV 30...
Страница 120: ...120 DR 6000 190 1100nm UV VIS 320 1100nm 190 360nm DR 6000 DR 6000...
Страница 121: ...121 DR 6000 DR 6000 USB US EU CN RFID 1 DR 6000 LINK2SC 10 40 C 50 104 F 80 15cm 138...
Страница 122: ...122 1 2 1 USB A 5 2 6 3 7 RFID 4 1 4 USB B 2 5 USB A 3 6...
Страница 123: ...123 1 122 2 100 240 V 50 60 Hz 3 On 122 1 2 1cm 3 1 4 3...
Страница 124: ...124 1 2 3 45 20 DR 6000 1 2 1 2 45 3 4 2 139...
Страница 125: ...125 1 2 1...
Страница 126: ...126 1 YZ_ USB USB 142 1 ABC abc 123 CE 2 HACH LANGE HACH HACH LANGE HACH LANGE HACH...
Страница 127: ...127 5000 ID ID 1 made to measure analysis HACH HACH LANGE Abs T 4 ID ID PC 2...
Страница 128: ...128 2 3 4 ID ID 5 XML Extensible Markup Language CSV Comma Separated Value USB DataLog...
Страница 130: ...130 6 200 1 2 3...
Страница 131: ...131 4 1 2 3 4 5 6 127 1 13mm ID 10 outlier elimination...
Страница 132: ...132 1 1 1 123 1...
Страница 133: ...133 1 2 2 4 3 127 190 1100nm Off On 132 1 1 5 1cm 7 127 127 3...
Страница 134: ...134 4 127 190 1100nm 132 1 1 5 1cm 7 127 132 4...
Страница 135: ...135 T Abs 5 USB A y 5...
Страница 136: ...136 1 2 3 4 1 1 5 1cm 7 127 127 5...
Страница 137: ...137 5 x 1 2 6...
Страница 138: ...138 y USB A 1 1 5 1cm 7 127 6...
Страница 139: ...139 3 5 3 5 LCW919 xx x...
Страница 140: ...140 999999 USB USB USB USB USB USB A USB...
Страница 141: ...141 USB USB USB USB FTP IP DHCP DHCP IP IP IP FTP FTP DR 6000 DR 6000...
Страница 142: ...142 A23778 A23792 LZV902 99 0002 0 EU YAA080 CH XLH051 UK XLH057 US 1801000 XLH069 A23772 LZV915 LZV886 USB LZV881...
Страница 172: ...172 CAN CSA C22 2 No 61010 1 1 2 II 2000 m 6560 ft...
Страница 173: ...173 RFID RFID DR 6000 RFID DR 6000 RFID DR 6000 RFID RFID 13 56 MHz 2005 8 12 2002 96 EC 30...
Страница 174: ...174 DR 6000 190 1100nm 320 1100 nm 190 360 nm DR 6000 DR 6000 DR 6000 DR 6000 USB...
Страница 175: ...175 RFID 1 DR 6000 LINK2SC 10 40 C 50 104 F 80 15 191 1 1 USB A 5 2 6 3 7 RFID 4...
Страница 176: ...176 2 1 176 2 100 240 V 50 60 Hz 3 On 176 1 4 USB B 2 5 USB A 3 6...
Страница 177: ...177 1 2 1 3 1 4 3 1 2 3 45 20 DR 6000 1 2 OK...
Страница 178: ...178 1 2 45 3 4 OK Main Menu 191 1 2 1...
Страница 179: ...179 1 YZ_ Cancel OK USB USB 193 Main Menu 1 ABC abc 123 CE...
Страница 180: ...180 5000 ID ID 1 Main Menu Recall Data 2 Main Menu Abs T System Check ID ID...
Страница 182: ...182 1 USB DR 6000 USB A DR 6000 2 Main Menu Recall Data 3 Data Log 4 Options PC Printer PC 5 6 OK...
Страница 184: ...184 4 5 Read Dilution 6 180 1 13 10 1 1 1 177 Barcode Programs 1 Information...
Страница 185: ...185 Dilution 1 Main Menu Single Wavelength 2 Parameter Setup Options 3 180 190 1100 nm 185...
Страница 186: ...186 Main Menu Multi Wavelength Parameter Setup Options Read 1 5 1 7 180 Instrument Setu p 180 3 4 180 190 1100 nm 185...
Страница 188: ...188 1 Wavelength Scan 2 View Table Graph View Table Track Peak Valley USB A y Read 1 5 1 7 180 Instrument Set up 180 5...
Страница 189: ...189 3 Zero Zeroing 4 5 Read Reading x Cursor Options Store...
Страница 190: ...190 1 Main Menu Time Course 2 Options 6 Scale Units y USB A Read 1 5 1 7 180 6...
Страница 191: ...191 3 5 3 5 LCW919 xx Start Again OK...
Страница 192: ...192 x 999999 U U U U U USB A...
Страница 194: ...194...
Страница 195: ...www hach com Hach Company 2012 2013 2018 All rights reserved Printed in Germany...