To ensure that the correct zero offset is applied and to eliminate any remaining TOC
contamination from the analyzer, a Service Zero cycle should be initiated using the SET SERVICE
ZERO function in the Service menu. When initiated, BioTector automatically runs a total of 10 zero
calibration cycles during the following 500 online measurements and automatically adjusts the
zero offset values to compensate for the removal of the contamination. It will therefore not be
necessary to revisit the BioTector after the startup or to repeat the Zero Calibration cycle
manually. See Section
for further details. (If ultra-pure water is used during the
Zero Calibration, ensure that container with at least 5 liters of water is connected to the analyzer. It
is recommended to use a sealed container with CO
2
absorber and HEPA air filter installed in line
on the vent/breathing tube to prevent the contamination of the water.)
_____
Download BioTector “All Data” in text format into the MMC/SD card using the SEND ALL DATA
function in Data Output menu to record all changes made in the system configuration. See Section
_____
Go to Start Stop menu and start the BioTector. When the BioTector is running online, carefully
observe the first two or three reactions and confirm that the CO
2
peaks are correct.
_____
Содержание BioTector B3500ul
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 W Watts Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...