8.3.5.2
Programmable Outputs
PWR BRD OUT 1-6
▪
POWER BOARD OUTPUT1-6 menus typically contain output
settings related to external devices. These relay outputs, located
on the Power and Input/Output Board, are system optional
features.
▪
POWER BOARD OUTPUT1-6 relays can be programmed to a
single output function or to multiple output functions as
described for OUTPUT 1-6 above. The programmed output
function or functions are mar
ked with an asterisk “*” sign in
these
menus as in the example for POWER BOARD OUTPUT 2,
which is programmed for STOP, FAULT, WARNING and NOTE
signals below.
▪
In the example below, POWER BOARD OUTPUT 2 is activated
when STOP or FAULT or WARNING or NOTE signal is
triggered.
▪
POWER BOARD OUTPUT 1 is a standard system relay
programmed to STOP, FAULT, WARNING and NOTE output
functions by default.
▪
In BioTector configuration data download (or in All Data
download), the programmed output functions for each output is
tabulated and marked with asterisk “*” signs for clarity
as
illustrated below.
PWR BRD OUT 2
DEFAULT STATE
N/E
STOP
*
FAULT
*
WARNING
*
NOTE
*
SAMPLER FILL
………………………
..
………………………
..
Содержание BioTector B3500ul
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 W Watts Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...