Page 115
SYSTEM PROGRAM 2 TC
SAMPLE IN 5p , P-V
▪
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
SYSTEM PROGRAM 2 VOC
▪
All system parameters for SYSTEM PROGRAM 2 VOC analysis
mode are covered in System Program 2 TC and System
Program 2 TIC + TOC above.
SYSTEM PROGRAM 3
SYSTEM PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3 150.0µgC/l
▪
RANGE CHANGE 2-3 defines the concentration level for the
BioTector to change its range automatically from range 2 to
range 3.
▪
The RANGE CHANGE 2-3 is typically the top concentration
point of TOC RANGE 2 in System Range Data menu by default.
RANGE CHANGE 3-2 120.0µgC/l
▪
RANGE CHANGE 3-2 specifies the concentration level for the
BioTector to change its range automatically from range 3 to
range 2.
▪
The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC RANGE 2 in System Range Data
menu by default.
RANGE CHANGE 3-1 40.0µgC/l
▪
RANGE CHANGE 3-1 defines the concentration level for the
BioTector to change its range automatically from range 3 to
range 1.
▪
The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC RANGE 1 in System Range Data
menu by default.
SAMPLE IN 3p P-V
▪
SAMPLE IN states the amount of sample (in pulses) to be
injected into the reactor at range 3.
TIC ACID 12p
▪
TIC ACID defines the amount of acid reagent (in pulses) to be
injected into the reactor at range 3.
BASE 70p
▪
BASE specifies the amount of base (in pulses) to be injected into
the reactor for the BASE OXIDATION phase at range 3.
BASE OXIDATION 90s
▪
BASE OXIDATION defines the oxidation time (90s by default),
which can be programmed independently for range 3.
▪
If the Base Oxidation time is different than the one programmed
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 3 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
TOC ACID 58p
▪
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 3.
Содержание BioTector B3500ul
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 W Watts Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...