Page 83
pH Test, Phase 5:
P H T E S T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < R A N G E , V A L V E 1 , S T R E A M 1
2 M O D E T I C + T O C
3 S T A R T T E S T
4 T A K E S A M P L E
5 C O N T I N U E T O N E X T P H A S E
6 S T O P T E S T
T I M E 0 s
P H A S E P A U S E D
M F C = 0 . 0 l / h C O 2 = 1 5 0 . 8 p p M
T E S T T O C p H . E X P E C T E D p H < 2 .
W H E N F I N I S H E D , S E L E C T N E X T A C T I O N
F R O M M E N U .
At this phase, the TOC acid is added to the solution in the reactor and mixed together. The system then
pauses to allow the pH to be tested. The user now has 3 options. Option 4 is used to take the sample as
before, but both options 5 and 6 below will end the test, as the TOC acid check is the last phase in the cycle.
5. Continue To Next Phase.
If this is selected, the program continues to the next phase, which is the
reactor purge phase.
6. Stop Test.
If this is selected, the program jumps to the reactor purge phase.
As the next phase is the reactor purge phase, the user is prompted to confirm that all tubes have been re-
connected before the BioTector starts this phase.
P H T E S T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < R A N G E , V A L V E 1 , S T R E A M 1
2 M O D E T I C + T O C
3 S T A R T T E S T
4 T A K E S A M P L E
5 C O N T I N U E T O N E X T P H A S E
6 S T O P T E S T
C O N F I R M A L L T U B E S R E - C O N N E C T E D
C O R R E C T L Y . P R E S S R I G H T A R R O W T O
C O N F I R M .
pH Test, Phase 6:
P H T E S T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < R A N G E , V A L V E 1 , S T R E A M 1
2 M O D E T I C + T O C
3 S T A R T T E S T
4 T A K E S A M P L E
5 C O N T I N U E T O N E X T P H A S E
6 S T O P T E S T
T I M E 0 s
P H A S E C O M P L E T E
M F C = 0 . 0 l / h C O 2 = 1 5 0 . 8 p p M
The pH test is complete. The BioTector will purge the reactor and the CO
2
analyzer. The user can either exit
the menu or start the pH test again.
Содержание BioTector B3500e
Страница 17: ...Page 17 Software Menu Diagram...
Страница 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Страница 61: ...Page 61...
Страница 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Страница 158: ...Page 158 Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...