2.1.3 Analysis Data Screen
B I O T E C T O R R U N N I N G 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
0 9 : 1 3 : 0 2 1 2 - 0 9 - 0 2 R E A C T I O N S T A R T
T I C & T O C S T R E A M 2 R E A C T I O N T Y P E
T O C R E A C T I O N P H A S E
1 R A N G E
2 6 6 s R E A C T I O N T I M E
3 6 0 s R E A C T I O N D U R A T I O N
R E A C T I O N R E S U L T T I C m g C / l T O C m g C / l
0 9 : 0 7 : 0 2 1 2 - 0 9 - 0 2 S 1 √ 1 3 0 . 0 5 4 0 . 0
0 9 : 0 1 : 0 2 1 2 - 0 9 - 0 2 S 2 √ 3 . 6 3 . 6
0 8 : 5 5 : 0 2 1 2 - 0 9 - 0 2 S 3 √ 7 . 2 7 . 2
0 8 : 4 9 : 0 2 1 2 - 0 9 - 0 2 S 4 x 1 0 . 7 1 0 . 7
0 8 : 4 3 : 0 2 1 2 - 0 9 - 0 2 S 5 x 1 4 . 3 1 4 . 3
0 8 : 3 7 : 0 2 1 2 - 0 9 - 0 2 C F 0 . 9 7 . 9
The Analysis Data screen is the default display screen on the BioTector for carbon (TIC, TOC and TC)
analysis. When the user moves through the various levels of the software menu, BioTector returns to this
screen automatically after 15 minutes if there is no further activity on the membrane keypad.
This screen gives information on:
- The Reaction Start time.
- The Reaction Type, for example a TIC & TOC reaction, TC reaction, Cleaning Reaction.
- The Reaction Phase, for example if the reaction is currently in the TIC, Base Oxidation, TOC phase.
- The operation Range (e.g. Range 1, 2 or 3) the BioTector is using to carry out its analysis.
- The Reaction Time, which is the elapsed time (seconds) since the analysis start.
- The Reaction Duration, which is the overall duration (seconds) of the analysis.
The Analysis Data screen also has an archive of the last 25 reactions. The most recent six reactions are
shown on the screen. In order to access the remaining reactions, use the DOWN or RIGHT keys to scroll
down, use the LEFT or UP keys to scroll up.
Each reaction record in the reaction archive contains:
- Start Time - reaction start time.
- Date - reaction date.
- Record Type, using the prefixes below:
S1
– reactions for stream 1.
M1
– reactions from manual sample stream 1.
√
– sample sensor detected the sample or there is no significant quantity of air bubbles in the
stream/manual grab sample lines.
x
– sample sensor detected no sample or there is significant quantity of air bubbles in the
stream/manual grab sample lines. See Sample Status in section
details.
RS
– remote standby reaction.
ZC
– zero calibration reaction.
ZK
– zero check reaction.
ZM
– manual zero adjust.
SC
– span calibration reaction.
SK
– span check reaction.
SM
– manual span factor adjust.
A1
– 24 hours average result from stream 1.
- Analysis Results – analysis results according to the analysis type (e.g. TIC, TOC in mgC/l).
Содержание BioTector B3500e
Страница 17: ...Page 17 Software Menu Diagram...
Страница 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Страница 61: ...Page 61...
Страница 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Страница 158: ...Page 158 Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...