![Hach BioTector B3500e Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/hach/biotector-b3500e/biotector-b3500e_user-manual_3202243054.webp)
Page 54
When the wiring of the system is completed, the power up of the system should be carried out in the order
below:
i)
While the external disconnection switches are powered off, power on the internal Main
Power Switch in the BioTector.
ii)
Close the BioTector electronics enclosure.
iii)
If installed, close the air compressor enclosure.
iv)
Switch on the external disconnection switches.
The power off of the BioTector should be carried out by switching off with the external disconnection switch
followed by the internal Main Power Switch. If installed, the power off of the air compressor should be carried
out by switching off with the external disconnection switch followed by compressor power switch.
Содержание BioTector B3500e
Страница 17: ...Page 17 Software Menu Diagram...
Страница 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Страница 61: ...Page 61...
Страница 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Страница 158: ...Page 158 Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...