Page 144
112_LO TEMP STANDBY
BioTector temperature has dropped
below 2°C for longer than 2 minutes.
BioTector has been automatically put
into standby mode and has stopped
analyzing.
When BioTector temperature increases
above 5°C for longer than 2 minutes,
BioTector continues normal operation.
Check ambient and BioTector
temperature.
Confirm BioTector is operating
within its specification
temperatures (5°C – 45°C).
113_TEMP SENSOR FAULT The difference between the
temperature readings recorded by
Micro Processor Unit (MPU)
temperature sensor (located on the
Motherboard) and the BioTector
temperature sensor (located on the O2-
Controller Board) has been outside the
±15°C temperature band.
Confirm that motherboard cover is
installed and the door of the
analyzer is closed.
Check ambient temperature,
which should be less than 45°C.
Confirm that the Analyzer is not
exposed to direct sun light.
If the problem is not related to
above, contact the distributor or
the manufacturer.
114_I/O WARNING
System has detected changes in the
Input/Output bus extender chips, which
have read/write control registers, during
the periodic checks carried out
automatically.
When system detects a
discrepancy between the
requested and the read
configuration registers values,
then all devices on SPI “Serial
Peripheral Interface” bus are reset
and re-initialized automatically.
Acknowledge the warning and
inform the distributor or the
manufacturer.
115_CO2 ANALYZER
WARN
Fault in CO
2
Analyzer.
Very dirty optics in CO
2
Analyzer.
Check CO
2
analyzer’s ppm CO
2
response in Simulate menu (see
Section
details).
Open CO
2
analyzer and clean the
optics.
Power down and power up the
BioTector.
If problem persists, check
outgoing 24V DC power to CO
2
analyzer on Motherboard at wires
N11 and N12.
For further tests, see information
sheet “T019. BioTector CO
2
Analyzer
Troubleshooting”, which is available in
the MMC/SD card shipped with the
BioTector.
Содержание BioTector B3500e
Страница 17: ...Page 17 Software Menu Diagram...
Страница 43: ...Page 43 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 45: ...Page 45 Figure 6 BioTector oxygen concentrator layout...
Страница 61: ...Page 61...
Страница 73: ...Page 73 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 152: ...Page 152 Section 11 System Replacement and Spare Parts...
Страница 158: ...Page 158 Z Zero ZC Zero calibration ZK Zero check ZM Manually input zero adjust ZS Zero and Span...
Страница 161: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...