42
sc1000 Operation
Figure 15
Measurement System—Removing the Flow Chamber
5.7.1.1 Buffer Solutions (pH)
The pH of buffer solutions are dependent on temperature. The nominal pH values are
referenced to a temperature of 25 °C (77 °F). Refer to
Table 4
for NIST standard buffer
solutions and DIN buffer values for a variety of temperatures. Always check the buffer
table for the value of the buffer in use at the operating temperature.
Table 4 NIST and DIN Buffers
Temp (°C)
NIST Buffer
DIN Buffer
Buffer 4.00
Buffer 6.88
Buffer 9.00
Buffer 4.00
Buffer 6.88
Buffer 9.00
0
4.01
6.984
9.464
4.05
7.13
9.24
5
4
6.951
9.395
4.04
7.07
9.16
10
4
6.923
9.332
4.02
7.05
9.11
15
4
6.9
9.276
4.01
7.02
9.05
20
4
6.881
9.225
4
7
9
25
4.01
6.865
9.18
4.01
6.98
8.95
30
4.01
6.853
9.139
4.01
6.98
8.91
35
4.02
6.844
9.102
4.01
6.96
8.88
40
4.03
6.838
9.068
4.01
6.85
8.85
45
4.04
6.834
9.038
4.01
6.9
8.82
50
4.06
6.833
9.01
4.01
6.95
8.79
55
—
6.833
8.985
4.01
6.95
8.76
Содержание 8362sc
Страница 2: ...Visit http www hach com...
Страница 6: ...Visit us at www hach com...
Страница 10: ...Visit us at www hach com...
Страница 14: ...14 Installation Figure 3 Measurement System Electrical Connection to Digital Gateway...
Страница 16: ...Visit us at www hach com...
Страница 31: ...31 sc100 Operation Figure 8 Overflow Sampling...
Страница 47: ...47 sc1000 Operation Figure 16 Overflow Sampling...
Страница 50: ...Visit us at www hach com...
Страница 52: ...Visit us at www hach com...
Страница 54: ...Visit us at www hach com...
Страница 58: ...Visit us at www hach com...