![Hach 59530-16 Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/hach/59530-16/59530-16_instruction-manual_3226394083.webp)
2—11
Error Codes,
continued
•
If the problem persists, contact a Service Center (
page 2—37
).
5.
E-6 Abs Error (User mode)
Indicates that the absorbance value is invalid, or indicates an attempt to make
a curve with less than two points.
•
Enter or measure the absorbance value again.
•
If the problem persists, contact a Service Center (
page 2—37
).
6.
E-7 Standard Value Error (User mode)
Standard concentration is equal to another standard concentration that is
already entered.
•
Enter the correct standard concentration.
•
If the problem persists, contact a Service Center (
page 2—37
).
7.
E-9 Flash Error
The meter is unable to save data.
•
If the problem persists, contact a Service Center (
page 2—37
).
Содержание 59530-16
Страница 10: ...1 10...
Страница 12: ...1 12...
Страница 15: ...1 15 Instrument Cap Cord continued Figure 1 Attaching the Instrument Cap Cord...
Страница 16: ...1 16...
Страница 52: ...1 52...
Страница 73: ...2 1 Section 2 Instrument Manual...
Страница 74: ...2 2...
Страница 79: ...2 7 Instrument Operation continued Figure 1 Battery Installation...
Страница 80: ...2 8...
Страница 100: ...2 28...
Страница 106: ...2 34...