Molybdenum, Molybdate, HR,
continued
Interferences
Interference studies were conducted by preparing a molybdenum standard solution
(2 mg/L Mo
6+
) as well as a solution of the potential interfering ion. When the
standard solution concentration changed by ± 5% with a given ion concentration,
the ion was considered an interference.
Large interferences are caused by some biocides common to cooling water tower
samples. Hach recommends testing the procedure on molybdenum standards in the
presence of the specific biocides in use to determine the feasibility of the ternary
complex chemistry.
.
Negative Interference:
Ion
Interferes at
(mg/L):
Ion
Interferes at
(mg/L):
Acrylates 790
Diethanoldithiocarba-mate
32
Alum
7
EDTA
1500
Aluminum
2
Ethylene Glycol
2%
(by volume)