![Hach 59530-16 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/hach/59530-16/59530-16_instruction-manual_3226394061.webp)
1—61
Ozone,
continued
5.
Repeat steps 1– 4 with cells labeled STD 2 and STD 3.
6.
Compare these measurements with previous measurements to verify the
instrument is performing consistently. (If these are the first measurements,
record them for comparison with later measurements.)
Note:
If the instrument is user-calibrated, initial standard measurements of the Spec
√
Standards will need to be performed again for the user calibration.
Summary of Method
The reagent formulation adjusts the sample pH to 2.5 after the ampule has filled.
The indigo reagent reacts immediately and quantitatively with ozone.
The blue color of indigo is bleached in proportion to the amount of ozone present
in the sample. Other reagents in the formulation prevent chlorine interference. No
transfer of sample is necessary in the procedure. Therefore, ozone loss due to
sampling is eliminated.