1—55
Ozone,
continued
4.
Fill one Indigo Ozone
Reagent AccuVac Ampul of
the appropriate range with
the sample and another
ampule with the blank.
Keep the tip immersed while
the ampule fills completely.
5.
Gently but quickly invert
both ampules several times
to mix. Wipe off any liquid
or fingerprints.
Note:
Do not shake the
ampules. Shaking will result
in a loss of ozone.
Note:
Part of the blue color
will be bleached if ozone is
present.
6.
Place the blank in the
cell holder.