![Hach 500sc SiO2 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/hach/500sc-sio2/500sc-sio2_operations_3154731009.webp)
1.
Select GRAB SAMPLE>GRAB SAMPLE OUT.
2.
Complete the instructions on the display.
3.
Put the grab sample out tube in a clean container. The grab sample tube is located in the lower
unit of the cabinet on the left side.
4.
Push down on the spigot (upper left corner of the lower compartment) for the grab sample tube to
dispense a grab sample.
Set up the system
The configuration settings can be changed in the SETUP SYSTEM menu or in the analyzer setup.
Refer to
on page 6.
1.
Select SETUP SYSTEM.
2.
Push
menu
and select SETUP SYSTEM.
3.
Select an option.
Option
Description
MEAS MODE
Changes the mode of the measurement cycle. Options: interval or continuous
(default). The continuous mode measures approximately every 9 minutes.
EDIT INTERVAL
(conditional)
Changes the interval time when MEAS MODE is set to interval. Options:
10–240 minutes (default = 15 minutes).
MEAS UNITS
Changes the measurement units that are shown on the display and in the data
log. Options: ppb (default), ppm, mg/L, µg/L.
SIGNAL AVERAGE
Selects the number of measurements used to calculate an average measurement
(1–5). This reduces variability in measurements (default = 1, no averaging).
EDIT ANALYZER
NAME
Changes the name that is shown on the top of the measurement screen
(maximum of 16 characters).
EDIT CHANNEL NAME
Changes the name of the sample source that is shown on the measurement
screen (maximum of 10 characters).
CONFIGURE
SEQUENCER (optional)
Starts or stops measurements for individual sample sources. Sets the
measurement order of the sample sources when there is more than one sample
source.
SAMPLE MISSING
Sets the operation that occurs if no sample is detected. Options: ON DELAY
(default) or OFF DELAY. ON DELAY: the instrument waits the remainder of the
measurement cycle, then goes to the next sample in the sequence. OFF DELAY:
the instrument waits 10 seconds, then goes to the next sample in the sequence.
SET DATE & TIME
Sets the analyzer time and date.
DISPLAY SETUP
Changes the language. Adjusts the order measurements are shown. Adjusts the
display contrast settings.
DISABLE REMINDERS
Stops scheduled maintenance alerts for individual components. Options: tubing,
stir bar, colorimeter cell, air filter, stir motor, air relief valve, pinch valve, air
compressor, reagent valves, sample valves, standard valves, colorimeter LED, fan
filter, air check valve.
MANAGE DEVICES
Installs or removes input modules. Refer to
on page 10 for
more information.
INSTRUMENT
INFORMATION
Shows the analyzer information. Refer to
on page 16.
CALCULATION
Set up variables, parameters, units and formulas for the analyzer. Refer to
on page 10.
SETUP OUTPUTS
Selects and configures the 4–20 mA setup, the relay setup and the error hold
mode. Refer to
on page 11 for more information.
English
9
Содержание 500sc SiO2
Страница 1: ...DOC023 97 80267 5500sc SiO2 07 2016 Edition 2 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...English 3 Fran ais 19 Espa ol 36 Portugu s 54 71 86 103 119 2...
Страница 86: ...86 94 89 98 91 SD 100 91 101 92 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 86...
Страница 88: ...2 6 1 2 Home 3 MEASUREMENT VALUE AUTO SCALE MANUALLY SCALE MANUALLY SCALE ppb DATE TIME RANGE 48 SD 4 SD 88...
Страница 89: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 89 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 4 YES 5 6 7 2 89...
Страница 103: ...103 110 106 115 107 SD 117 108 117 109 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 103...
Страница 104: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 104...
Страница 119: ...119 126 122 130 123 SD 132 124 133 124 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE MENU 3 8 Back 4 9 Menu 5 Enter 2 3 119...
Страница 120: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 120...
Страница 134: ......
Страница 135: ......