Option
Description
DERIVATIVE
Sets a value that adjusts for vacillation of the process. The majority of applications can be
controlled without the use of the derivative setting.
TRANSIT TIME
Sets the value to stop the PID control for a selected period of time when the sample moves
from the control pump to the measurement sensor.
5.
Select SET FUNCTION>LOGARITHMIC and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET 50% VALUE
Sets the value corresponding to 50% of the process variable range.
SET HIGH VALUE
Sets the high endpoint (upper value) of the process variable range.
6.
Select SET FUNCTION>BILINEAR and then select the applicable options in the ACTIVATION
menu.
Option
Description
SET LOW VALUE
Sets the low endpoint of the process variable range.
SET HIGH VALUE
Sets the high endpoint (upper value) of the process variable range.
SET KNEE POINT VALUE
Sets the value at which the process variable range divides into another linear
segment.
SET KNEE POINT CURRENT
Sets the value of the current at the knee point value.
Relay setup
1.
Push
menu
and select SETUP SYSTEM>SETUP OUTPUTS>RELAY SETUP.
2.
Select the relay.
3.
Select an option.
Option
Description
ACTIVATION
The menu list items change with the selected function. Refer to
on page 12 for more information.
SELECT SOURCE
Selects the output. Options: None (if the relay is not configured), the analyzer name or
calculation (if a calculation formula has been configured). Refer to
on page 10.
SET FUNCTION
Selects a function. ALARM—The relay starts when the upper or lower alarm value is
triggered. FEEDER CONTROL—The relay shows if a process value is larger or falls
below a setpoint. EVENT CONTROL—The relay toggles if a process value reaches an
upper or lower limit. SCHEDULER—The relay switches at certain times independently of
any process value. WARNING—The relay shows warning and error conditions in
probes. PROCESS EVENT—The relay switches when the analyzer does a specified
operation.
SET TRANSFER
Selects active or inactive.
FAIL SAFE
Selects yes or no.
Relay activation options
1.
Push
menu
and select SETUP SYSTEM>SETUP OUTPUTS>RELAY SETUP.
2.
Select the applicable relay.
12
English
Содержание 500sc SiO2
Страница 1: ...DOC023 97 80267 5500sc SiO2 07 2016 Edition 2 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...English 3 Fran ais 19 Espa ol 36 Portugu s 54 71 86 103 119 2...
Страница 86: ...86 94 89 98 91 SD 100 91 101 92 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 86...
Страница 88: ...2 6 1 2 Home 3 MEASUREMENT VALUE AUTO SCALE MANUALLY SCALE MANUALLY SCALE ppb DATE TIME RANGE 48 SD 4 SD 88...
Страница 89: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 89 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 4 YES 5 6 7 2 89...
Страница 103: ...103 110 106 115 107 SD 117 108 117 109 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 103...
Страница 104: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 104...
Страница 119: ...119 126 122 130 123 SD 132 124 133 124 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE MENU 3 8 Back 4 9 Menu 5 Enter 2 3 119...
Страница 120: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 120...
Страница 134: ......
Страница 135: ......