![Hach 500sc SiO2 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/hach/500sc-sio2/500sc-sio2_operations_3154731010.webp)
Option
Description
SETUP NETWORK
(conditional)
Shows only if a network card is installed. The network card support is for Modbus,
Profibus and HART.
SECURITY SETUP
Enables or disables the pass code (default = HACH55).
AIR PURGE
Enables the air purge for use with an external air supply. Options: on or off
(default). Off: an external air supply is not used. The vent fan is enabled and the
air filter is installed. On: an external air supply is connected to the instrument. The
vent fan is disabled. The air filter is replaced with a fan filter plug. To use this
feature, make sure that the fan filter plug is installed. Refer to the documentation
that is supplied with the air purge kit.
RESET DEFAULTS
Sets the configuration to the factory defaults.
Manage devices
Install or remove input modules.
1.
Push
menu
and select MANAGE DEVICES.
2.
Select an option.
Option
Description
SCAN FOR DEVICES
The system shows the connected devices.
Note: If no device is connected, the system returns to the main measurement
screen.
DELETE DEVICE
Removes the device when the device is no longer connected.
Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1.
Push
menu
and select CALCULATION.
2.
Select an option.
Option
Description
SET VARIABLE X
Selects the sensor referring to the variable X.
SET PARAMETER X
Selects the parameter referring to the variable X.
SET VARIABLE Y
Selects the sensor referring to the variable Y.
SET PARAMETER Y
Selects the parameter referring to the variable Y.
SET FORMULA
Selects the calculation formula to complete. Options: None, X-Y, X+Y, X/Y, [X/Y]%, [X
+Y]/2, X*Y, [X-Y]%/X
DISPLAY FORMAT
Selects the number of decimal places shown in a calculation result. Options: Auto,
XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
SET UNITS
Enters the unit name (maximum of 5 characters).
SET PARAMETER
Enters the measurement name (maximum of 5 characters).
10
English
Содержание 500sc SiO2
Страница 1: ...DOC023 97 80267 5500sc SiO2 07 2016 Edition 2 Operations Fonctionnement Operaciones Opera es...
Страница 2: ...English 3 Fran ais 19 Espa ol 36 Portugu s 54 71 86 103 119 2...
Страница 86: ...86 94 89 98 91 SD 100 91 101 92 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 86...
Страница 88: ...2 6 1 2 Home 3 MEASUREMENT VALUE AUTO SCALE MANUALLY SCALE MANUALLY SCALE ppb DATE TIME RANGE 48 SD 4 SD 88...
Страница 89: ...4 SD 1 3 LED 2 2 4 SD 1 1 1 2 3 4 89 4 1 LANGUAGE 2 DATE FORMAT 3 DATE TIME 4 YES 5 6 7 2 89...
Страница 103: ...103 110 106 115 107 SD 117 108 117 109 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE 3 8 Back 4 Home 9 Menu 5 Enter 2 3 103...
Страница 104: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 104...
Страница 119: ...119 126 122 130 123 SD 132 124 133 124 1 1 1 6 Diag DIAG TEST 2 7 Cal CALIBRATE MENU 3 8 Back 4 9 Menu 5 Enter 2 3 119...
Страница 120: ...2 1 6 2 7 3 8 4 9 5 SD SD 10 3 1 3 Rx Sx 1 2 PROGNOSYS 4 PROGNOSYS PROGNOSYS 0 100 45 75 10 45 50 75 10 50 1 120...
Страница 134: ......
Страница 135: ......