2
A/M button disabled, set point R/W only.
Lockout Menu
17.
Push INFINITY.
The sample heater returns to normal operations.
Sample heater operation menu
The upper display window shows the process temperature. The lower display window shows the
temperature setpoint. The 1 indicator lamp is on when the heater is on. No other display indicators
show.
If necessary, use the up and down arrow keys to change a value shown in the display, then move to
the next step.
1.
To go to the Lockout menu, push and hold the up and down arrow keys for approximately
3 seconds. Change the upper display to 5.
5
Operations menu and setup menu with full R/W access
Lockout menu
2.
Push INFINITY.
The sample heater enters the operation menu.
3.
Push
ADVANCE
.
Make sure that this is set to
no
. (To initially tune the
heater, select
YES
. Refer to
Autotune
4.
Push
ADVANCE
.
Make sure that this is set to
Auto
Control Mode Active
5.
Push
ADVANCE
.
Set to 9
Heat Proportional Band
6.
Push
ADVANCE
.
Set to 109
Time Integral
7.
Push
ADVANCE
.
English
17
Содержание 48685
Страница 2: ...DOC023 97 80196 Sample Heater 02 2012 Edition 3 User Manual Manuel d utilisation...
Страница 3: ...English 3 Fran ais 20 2...
Страница 10: ...Figure 5 Sample heater front view English 9...
Страница 38: ......