Color codes for the power cord
on page 12.
Color codes for the power cord
shows the color codes for the power cord.
Table 2 Color codes for the power cord
Conductor Type
North American
IEC
Line conductor (Line 1 (Hot))
Black
Brown
Neutral conductor (Line 2 (Neutral))
White
Blue
Ground conductor
Green
Green with yellow stripe
Terminal block connections
Connect the ground line to the heater block terminal that shows the ground symbol (
).
Connect the hot and neutral power supply lines to TB1 (
). The terminal block accepts up to
12 gauge wire. The manufacturer recommends 16–12 AWG wire based on installation requirements.
The wiring insulation must be rated to 90 °C minimum.
shows the interior of the sample heater with connections.
Figure 7 Power connections at the J3 terminal block
1
Line 1 (Hot)
3
Ground
2
Line 2 (Neutral)
4
Top of instrument
12
English