SECTION 3,
continued
16.
When all the digits
show the appropriate
value, press
READ
.
The instrument will
count from 60 to 0
(67 to 0 if signal average
is on), measure the
turbidity and store the
value. The instrument
will increment back to
S0
. Remove the sample
cell from the cell
compartment.
17.
Press:
CAL
.
The instrument will store
the new calibration data
and return the instrument
to the measurement
mode. It will use the new
calibration to calculate
turbidity for subsequent
measurements.
Note:
Pressing
CAL
completes the calculation
of the calibration
coefficients. If calibration
errors occurred during
calibration, error messages
will appear after
CAL
is
pressed. If
E 1
or
E 2
appear, check the standard
preparation and review the
calibration; repeat the
calibration if necessary.
If
CAL?
appears, an error
may have occurred during
calibration. If
CAL?
is
flashing, the instrument
is using the default
calibration.
READ