1.2 Accessories
Accessories supplied with the turbidimeter include nine sample
cells; three Gelex
®
Secondary Standards; one sealed vial each of:
<0.1-NTU, 20-NTU, 100-NTU, and 800-NTU StablCal
®
Stabilized
Formazin Standards; 4 AA alkaline batteries; 15 mL of silicone oil;
oiling cloth; carrying case; instrument manual; and quick reference card.
1.3 Principle of Operation
The Model 2100P Portable Turbidimeter operates on the nephelometric
principle of turbidity measurement. This instrument meets the design
criteria specified by the United States Environmental Protection
Agency, Method 180.1.
The optical system* (
Figure 2
) includes a tungsten-filament lamp, a 90°
detector to monitor scattered light and a transmitted light detector. The
instrument's microprocessor calculates the ratio of the signals from the
90
°
and transmitted light detectors. This ratio technique corrects for
interferences from color and/or light absorbing materials (such as
activated carbon) and compensates for fluctuations in lamp intensity,
providing long-term calibration stability. The optical design also
minimizes stray light, increasing measurement accuracy.
Figure 2
Ratio Optical System
* Patent number 4,198,161; other patents pending.
Lamp
Transmitted
Light
Detector
90°
Detector
Sample
Cell
Lens