SECTION 3,
continued
10.
The display will
show the
S2
icon (with
the
2
flashing) and
100 NTU
or the value
of the S2 standard for the
previous calibration.
11.
Edit the standard
concentration by
pressing
→
. The
2
will
stop flashing and the left
digit in the display will
flash. Press
↑
to scroll the
digit up to the
appropriate number.
Press
→
again to move
the cursor to the next
digit and edit it in the
same manner.
12.
When all the digits
show the appropriate
value, press
READ
.
The instrument will
count from 60 to 0
(67 to 0 if signal average
is on), measure the
turbidity and store the
value. Remove the
sample cell from the
cell compartment.
13.
Thoroughly mix the
700 to 900 NTU
standard, then fill a clean
sample cell to the line
with the standard. Insert
the cell into the cell
compartment.
14.
The display will
show the
S3
icon (with
the
3
flashing) and
800 NTU
or the value
of the S3 standard for the
previous
15.
Edit the standard
concentration by
pressing
→.
The
3
will
stop flashing and the left
digit in the display will
flash. Press
↑
to scroll the
digit up to the
appropriate number.
Press
→
again to move
the cursor to the next
digit and edit it in the
same manner.
NTU
CAL
READ
NTU
CAL