32
Troubleshooting
3.
Press
4 (RET)
to return to the main screen.
4.
Program the date and time and any other function that may have been set to original
default values.
6.3 Troubleshoot the pump
If the vacuum pump does not operate or exhibits any questionable symptoms such as
unusual sound or vibration, it may be removed for maintenance to restore it to optimum
operation.
6.3.1 Disassemble and clean the pump
Prerequisites:
•
1.27 mm or 1.5 mm Allen wrench
To disassemble and clean the pump:
1.
Open the counter as described in steps 1 to 3 of
2.
Remove the pump from the chassis clip.
3.
Unplug the pump electrical connection.
4.
Disconnect the tubing from the pump.
Note: Make note of the tubing that is attached to the Vacuum (V) fitting and the Pressure (P)
fitting. This information is required for the assembly of the pump.
Figure 26 Default mode screen
Figure 27 Pump