5
Section 1
Specifications
Specifications are subject to change without notice.
General
Dimension (W x H x L)
101.60 mm x 58.42 mm x 165.10 mm (4.0" x 2.3" x 6.5")
Weight
1 kg (2.13 lb)
Maximum count displayed
9,999,999
Sample flow rate
0.1 cfm
Number of size ranges
2 (second size range can be selected at the front panel)
Particle size ranges
MET ONE 227A—0.5 micron and larger (channel 1); 0.7 micron, 1 micron, 3
micron or 5 micron (channel 2)
MET ONE 227B—0.3 micron and larger (channel 1); 0.5 micron, 1 micron, 3
micron or 5 micron (channel 2)
Coincidence error
Less than 5% at 2,000,000 particles/cubic foot
Power
+5 W, +10 V (approximately) (+6 V with battery pack)
Light source
Laser diode
Sample/hold times
1 second to 24 hours
Count alarms
1 to 9,999,999 counts
Data storage
200 samples, rotating buffer
Count cycles
Up to 100 in Auto or Average mode
Locations
Up to 250 (number appears on the printout)
Output
RS232C/RS485 for computer
Pump type
Carbon vane, 0.1 cfm
Recharging time
Up to 90% in 10 hours, 100% in 12 hours
Battery type
Rechargeable NiMH, 3500 mAH
Operating time
8 hours, typical
Environment
Operating temperature
12 to 29 °C (55 to 84 °F)
Operating humidity
10 to 85% relative humidity, non-condensing
Storage temperature
-40 to 71 °C (-40 to 160 °F)
Storage humidity
Up to 98% relative humidity, non-condensing