6.2 Adjusting the image upright position (Second observer

tube long)

The image visible through the eyepiece is turned about the visual axis by adjusting

the viewing height of the second observer. The image is uprighted by adjusting the

twist grip (8).

7 Decommissioning

Decommissioning is not required. However, we recommend dismounting the second

observer tube (long) after use in order to prevent accidental impacts.

8 Technical data

8.1 Second observer tube short

Focal length f

132 mm

Image size and frame Are identical in the binocular tube and second observer

tube

Length of the tube

~150 mm

Stop for tonometer

Can be mounted / dismantled as required

8.2 Second observer tube long

Focal length f

132 mm

Image size and frame Are identical in the binocular tube and second observer

tube

Length of the tube

~410 mm

Stop for tonometer

Can be mounted / dismantled as required

9 Maintenance

WARNING!

• Do not modify this device without authorization of the manufacturer.

Installation and repairs may only be performed by trained

specialists.

• Contact your Haag-Streit representative for installation, repairs and

modification work on the system. The contact details are available at

www.haag-streit.com.

• Only original Haag-Streit spare parts may be used.

9.1 Device inspection

The following procedure should be used to check that the second observer tubes

are working correctly:

• Insert the test rod into the radial movement bearing, whilst at the same time

aligning the surface to the microscope at a right angle.

• Set the slit length to 8 or 14 mm.

• Set the illumination intensity to 50%.

• Set the magnification in the microscope to max.

• Set the eyepieces in such a way that the test rod is in sharp focus. In doing so,

turn the ocular from the (+) to the (-) side.

• For all magnifications, the structure of the test rod must be shown in sharp

focus.

• In the second observer tubes, the same area must also be in sharp focus at all

magnifications

9.2 Servicing

To guarantee a long service life, the device must be cleaned weekly as described

and protected with the dust cover when not in use. We recommend having the

device inspected once a year by an authorized service technician.

Содержание Second observer tube

Страница 1: ...EUTSCH ENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 INSTRUCTIONS FOR USE Second observer tube 4 Edition 2021 08 ...

Страница 2: ...ysician or licensed practitioner Contents 1 Safety 3 1 1 Comments on these instructions for use 3 1 2 Ambient conditions 3 1 3 Shipment and unpacking 3 1 4 Installation warnings 3 1 5 Operation environment 3 1 6 Disinfection 4 1 7 Warranty and product liability 4 1 8 Reporting obligation 4 1 9 Description of symbols 4 2 Purpose of use 4 3 Introduction 5 3 1 Overview 5 4 Device assembly installatio...

Страница 3: ...is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport company Make a report of any damaged parts This report must be signed by you and by the representative of the transport company Leave the device in the packaging for a few hours before unpacking it condensation Check the device for damage after it is unpacked R...

Страница 4: ... such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed or implied including implied warranties of merchantability or fitness for a particular use Haag Streit expressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller Fo...

Страница 5: ...rm locking screw 14 Cover 4 Device assembly installation NOTE If the tube is mounted on the same side as the tonometer the tonometer stop 5 must be fixed on the end of the tube If the tube is left on the slit lamp permanently the selectable relief springs 10 must be reset See instruction manual for the BQ 900 slit lamp Tighten the control knob 12 on the cross slide to block the microscope arm and ...

Страница 6: ...e varied so that the second observer can look into the eyepiece at a freely adjustable inclination angle FORBIDDEN The viewing height should not be adjusted during the examination as the patient or person carrying out the examination could be injured To adjust the viewing height the release button 9 must be released first of all To do this the elbow at the side of the beam splitter must be held so...

Страница 7: ... modify this device without authorization of the manufacturer Installation and repairs may only be performed by trained specialists Contact your Haag Streit representative for installation repairs and modification work on the system The contact details are available at www haag streit com Only original Haag Streit spare parts may be used 9 1 Device inspection The following procedure should be used...

Страница 8: ...ive disinfectants or cleaning liquids e g hydrogen peroxide will damage the finish and coating of the device Do not use sprays Observe the manufacturer s safety instructions Do not use any cloths that drip Wring out any soaked cloths before use when necessary Ensure that no liquid penetrates the device Comply with the stipulated exposure time Clean optical surfaces after disinfection with a very s...

Страница 9: ...ALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 Should you have any further questions please contact your Haag Streit representative at Wiki Revision ID 117729 ...

Отзывы: