3 Introduction

The observation in the second observer tube is monocular and takes place in the

right or left-hand side of the beam path of the stereo-microscope depending on the

slit lamp connection.

3.1 Overview

Required accessories for the slit lamp BQ 900:

1. Beam splitter 50/50 Haag-Streit

part number 1400308

2.

Second observer tube short

(1400301)

3. Eyepiece

4. CE label / Haag-Streit part number

(back) type designation

5. Tonometer stop

6. Union nut

7.

Second observer tube long

(1400034)

8. Twist grip for image upright

position

9. Release button

10. Selectable relief springs

11. Bayonet joint

12. Joy stick base locking screw

13. Microscope arm locking screw

14. Cover

4 Device assembly / installation

NOTE!

• If the tube is mounted on the same side as the tonometer, the

tonometer stop (5) must be fixed on the end of the tube.

• If the tube is left on the slit lamp permanently, the selectable relief

springs (10) must be reset (See instruction manual for the BQ 900

slit lamp)

• Tighten the control knob (12) on the cross slide to block the microscope arm and

the screw to block (13) horizontal movements.

• Loosening the bayonet joint (11) on the microscope (caution, hold the

microscope tight!) enables the stereo-microscope to be dismantled and the

beam splitter to be mounted.

• Carefully mount the stereo-microscope on the beam splitter with the bayonet

joint.

• Loosen the lid mounted on the side of the beam splitter (14).

• Hold the second observer tube in accordance with the figure. Align the groove

on the end of the tube vertically to the wedges of the beam splitter connecting

sleeve.

• Push the tube into the stop and fix it in place with the union nut (6).

Содержание Second observer tube

Страница 1: ...EUTSCH ENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 INSTRUCTIONS FOR USE Second observer tube 4 Edition 2021 08 ...

Страница 2: ...ysician or licensed practitioner Contents 1 Safety 3 1 1 Comments on these instructions for use 3 1 2 Ambient conditions 3 1 3 Shipment and unpacking 3 1 4 Installation warnings 3 1 5 Operation environment 3 1 6 Disinfection 4 1 7 Warranty and product liability 4 1 8 Reporting obligation 4 1 9 Description of symbols 4 2 Purpose of use 4 3 Introduction 5 3 1 Overview 5 4 Device assembly installatio...

Страница 3: ...is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport company Make a report of any damaged parts This report must be signed by you and by the representative of the transport company Leave the device in the packaging for a few hours before unpacking it condensation Check the device for damage after it is unpacked R...

Страница 4: ... such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed or implied including implied warranties of merchantability or fitness for a particular use Haag Streit expressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller Fo...

Страница 5: ...rm locking screw 14 Cover 4 Device assembly installation NOTE If the tube is mounted on the same side as the tonometer the tonometer stop 5 must be fixed on the end of the tube If the tube is left on the slit lamp permanently the selectable relief springs 10 must be reset See instruction manual for the BQ 900 slit lamp Tighten the control knob 12 on the cross slide to block the microscope arm and ...

Страница 6: ...e varied so that the second observer can look into the eyepiece at a freely adjustable inclination angle FORBIDDEN The viewing height should not be adjusted during the examination as the patient or person carrying out the examination could be injured To adjust the viewing height the release button 9 must be released first of all To do this the elbow at the side of the beam splitter must be held so...

Страница 7: ... modify this device without authorization of the manufacturer Installation and repairs may only be performed by trained specialists Contact your Haag Streit representative for installation repairs and modification work on the system The contact details are available at www haag streit com Only original Haag Streit spare parts may be used 9 1 Device inspection The following procedure should be used...

Страница 8: ...ive disinfectants or cleaning liquids e g hydrogen peroxide will damage the finish and coating of the device Do not use sprays Observe the manufacturer s safety instructions Do not use any cloths that drip Wring out any soaked cloths before use when necessary Ensure that no liquid penetrates the device Comply with the stipulated exposure time Clean optical surfaces after disinfection with a very s...

Страница 9: ...ALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 Should you have any further questions please contact your Haag Streit representative at Wiki Revision ID 117729 ...

Отзывы: