© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220394‑04040 – 2021 – 08

1 Safety

DANGER!

Failure to comply with these instructions may result in material damage

or pose a danger to patients or users.

WARNING!

These warnings must absolutely be complied with to guarantee safe

operation of the product and to avoid any danger to users and to

patients.

NOTE!

Important information, please read carefully.

1.1 Comments on these instructions for use

NOTE!

In these instructions for use the point is used as decimal separator.

1.2 Ambient conditions

Temperature

−40 °C ... +70 °C

Air pressure

500 hPa ... 1060 hPa

Transport

Relative humidity

10 % ... 95 %

Temperature

−10 °C ... +55 °C

Air pressure

700 hPa ... 1060 hPa

Storage

Relative humidity

10 % ... 95 %

Temperature

+10 °C ... +35 °C

Air pressure

800 hPa ... 1060 hPa

Use

Relative humidity

30 % ... 90 %

1.3 Shipment and unpacking

• Before unpacking the device, check whether the packaging shows traces of

improper handling or damage. If this is the case, notify the transport company

that delivered the goods to you.

• Unpack the device together with a representative of the transport company.

Make a report of any damaged parts. This report must be signed by you and by

the representative of the transport company.

• Leave the device in the packaging for a few hours before unpacking it

(condensation).

• Check the device for damage after it is unpacked.

• Return defective devices in the appropriate packaging.

• Store packaging material carefully so that it can be used for potential returns or

when moving.

• Check that the connection parts of the accessories are in the correct position

(screw connections, quick-release fasteners).

1.4 Installation warnings

WARNING!

• Do not modify this device without authorization of the manufacturer.

Installation and repairs may only be performed by trained

specialists.

• Only original Haag-Streit spare parts may be used.

1.5 Operation, environment

WARNING!

• Due to the risk of injury, it must not be possible for passing persons

to bump against the second observer tube during an examination.

• The second observer tube must be mounted where its contours

which protrude over the table top will not impede passing persons.

• Avoid moisture.

Содержание Second observer tube

Страница 1: ...EUTSCH ENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 INSTRUCTIONS FOR USE Second observer tube 4 Edition 2021 08 ...

Страница 2: ...ysician or licensed practitioner Contents 1 Safety 3 1 1 Comments on these instructions for use 3 1 2 Ambient conditions 3 1 3 Shipment and unpacking 3 1 4 Installation warnings 3 1 5 Operation environment 3 1 6 Disinfection 4 1 7 Warranty and product liability 4 1 8 Reporting obligation 4 1 9 Description of symbols 4 2 Purpose of use 4 3 Introduction 5 3 1 Overview 5 4 Device assembly installatio...

Страница 3: ...is the case notify the transport company that delivered the goods to you Unpack the device together with a representative of the transport company Make a report of any damaged parts This report must be signed by you and by the representative of the transport company Leave the device in the packaging for a few hours before unpacking it condensation Check the device for damage after it is unpacked R...

Страница 4: ... such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed or implied including implied warranties of merchantability or fitness for a particular use Haag Streit expressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller Fo...

Страница 5: ...rm locking screw 14 Cover 4 Device assembly installation NOTE If the tube is mounted on the same side as the tonometer the tonometer stop 5 must be fixed on the end of the tube If the tube is left on the slit lamp permanently the selectable relief springs 10 must be reset See instruction manual for the BQ 900 slit lamp Tighten the control knob 12 on the cross slide to block the microscope arm and ...

Страница 6: ...e varied so that the second observer can look into the eyepiece at a freely adjustable inclination angle FORBIDDEN The viewing height should not be adjusted during the examination as the patient or person carrying out the examination could be injured To adjust the viewing height the release button 9 must be released first of all To do this the elbow at the side of the beam splitter must be held so...

Страница 7: ... modify this device without authorization of the manufacturer Installation and repairs may only be performed by trained specialists Contact your Haag Streit representative for installation repairs and modification work on the system The contact details are available at www haag streit com Only original Haag Streit spare parts may be used 9 1 Device inspection The following procedure should be used...

Страница 8: ...ive disinfectants or cleaning liquids e g hydrogen peroxide will damage the finish and coating of the device Do not use sprays Observe the manufacturer s safety instructions Do not use any cloths that drip Wring out any soaked cloths before use when necessary Ensure that no liquid penetrates the device Comply with the stipulated exposure time Clean optical surfaces after disinfection with a very s...

Страница 9: ...ALIANO ESPAÑOL NEDERLANDS PORTUGUÊS SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc no 1500 7220394 04040 2021 08 Should you have any further questions please contact your Haag Streit representative at Wiki Revision ID 117729 ...

Отзывы: